TY - JOUR T1 - AUtomatic detection and recording of cardiac arrhythmias AU - Haber E Y1 - 1959/08/08 N1 - 10.1001/jama.1959.03010150026007 JO - Journal of the American Medical Association SP - 1782 EP - 1785 VL - 170 IS - 15 N2 - The cardiac monitor here described is a portable instrument used in connection with a standard electrocardiograph. From this source the monitor receives impulses which are amplified and interpreted, actuate the recording mechanism, cause an alarm to sound, and can, optionally, start up an electrical cardiac stimulator. Five cases illustrate its operation in patients with episodes of ventricular asystole, asymptomatic changes in intraventricular conduction, a paroxysm of bigeminal rhythm, an episode of extreme nodal bradycardia, and bouts of asymptomatic paroxysmal ventricular tachycardia. In a sixth patient the instrument recorded a sudden change from a sinus rhythm at 65 to a nodal rhythm at 80 beats per minute. This sort of continuous monitoring has evident advantages over intermittent observation. SN - 0002-9955 M3 - doi: 10.1001/jama.1959.03010150026007 UR - http://dx.doi.org/10.1001/jama.1959.03010150026007 ER -