Original Contribution |

Management of Recalled Pacemakers and Implantable Cardioverter-Defibrillators:  A Decision Analysis Model FREE

Mitesh S. Amin, MD; David B. Matchar, MD; Mark A. Wood, MD; Kenneth A. Ellenbogen, MD
[+] Author Affiliations

Author Affiliations: Department of Internal Medicine (Dr Amin) and Division of Cardiology (Drs Wood and Ellenbogen), Virginia Commonwealth University Medical Center, Richmond; Department of Medicine and Center for Clinical Health Policy Research, Duke University, and Durham Department of Veterans Affairs Medicine Center, Durham, NC (Dr Matchar).

More Author Information
JAMA. 2006;296(4):412-420. doi:10.1001/jama.296.4.412.
Text Size: A A A
Published online

Context Limited information exists to direct clinical management after an implantable device has been put under advisory. A better understanding of the risks and benefits of device replacement compared with continued clinical follow-up would be helpful to clinicians.

Objective Using the tools of decision analysis, to determine the best management approach (immediate device replacement vs continued monitoring) in the setting of a device advisory.

Design A decision model was constructed to evaluate the risks and benefits associated with immediate device replacement compared with continued monitoring.

Main Outcome Measures Variables considered included indications for device implantation, anticipated course following device failure, device failure rates from the advisory ranging from 0.0001% to 1.0% per year, and device replacement mortality rates ranging from 0.10% to 1.00% per procedure. Device replacement was preferred to continued follow-up when replacement led to greater patient survival.

Results The decision to replace a recalled device depends primarily on the advisory's estimated device failure rate and the likely effects of device failure on mortality. Procedural mortality is an important secondary factor, while patient age and remaining generator life have the least influence on the decision. For pacemaker-dependent patients, advisory device failure rates exceeding 0.3% warrant device replacement in most situations. In patients with implantable cardioverter-defibrillators for primary or secondary prevention, a failure rate associated with an advisory of 3.0% is needed to favor replacement in most cases, decreasing to close to 1.0% as procedural mortality rates decrease to 0.1% or risk of fatal arrhythmias increase to near 20% per year. In cases of pacemaker implantation for non–life-threatening situations (eg, carotid sinus hypersensitivity), most device advisories do not warrant device replacement.

Conclusions The decision to replace a device under advisory is determined primarily by the incidence of device malfunction and the likely effects of device failure. This analysis provides a framework for managing recalled devices in the context of device, patient, and institutional characteristics.

Figures in this Article

Currently, there are approximately 2 million patients with implantable cardioverter-defibrillators (ICDs) and pacemakers worldwide.1 These devices are unique because they are often implanted for life-threatening conditions—pacemakers for patients with high-grade heart block and ICDs to provide therapy for patients at risk of sudden cardiac death.

An increasing number of device recalls and advisories has led to heightened awareness of potential device failure.13 During 2005, US Food and Drug Administration advisories affected more than 200 000 devices.1 Advisories typically arise from unanticipated device failures that are identified after product release and widespread clinical use.

Despite the increased awareness of device malfunctions, there is a lack of consensus about optimal clinical management in situations involving recalled devices. Formal recommendations are few, with decisions mainly left to physician judgment.4 A recent survey reported that physicians were replacing more than 30% of the devices under advisory, although the percentage varied widely from 0% to 100% among individual physicians.5 Device replacement, however, carries risks of complications, including death.

In the face of multiple competing issues and uncertainty, decision modeling can provide useful guidance.6,7 Modeling offers quantitative estimates of net benefit and permits examination of the impact of assumptions on preferred outcomes.6,7 We developed a decision model using a Markov structure to examine the implications of immediate device replacement vs continued monitoring of devices under advisory, considering factors relevant to clinical decision making.

Our approach directly addresses the competing issues of recalled devices: the risk of death due to device replacement vs the risk of death due to device malfunction. To examine whether to replace a device after an advisory, we focus on the indications for device implantation and the associated potential outcomes of device failure. Table 1 shows 6 scenarios grouped into 3 different categories, which are used throughout this article.

Table Graphic Jump LocationTable 1. Scenarios of Device Failure Characteristics and Risk of Events
Decision Model

The question of whether to immediately replace a recalled device or to follow the patient until scheduled replacement (continued monitoring) is represented as a Markov model (Figure 1). The primary outcome measure of the model is life expectancy for each strategy (immediate replacement vs continued monitoring).

Figure 1. Simplified Diagram of the Markov Decision Analysis Model
Graphic Jump Location

Squares indicate decision nodes, choices facing the decision maker. Circles indicate chance nodes, events that have multiple possible outcomes and are not in the decision maker’s control. The square at the left indicates 2 initial management options, immediate device replacement and continued monitoring. After the initial decision, each option enters a Markov process, which leads to 1 of many health states. The initial path within the Markov process is forced (bold), device replacement state for the immediate replacement decision and device under advisory state for the continued monitoring decision. The Markov cycle represents a single 1-month cycle of the model (ie, 1 month of a patient's life), and the end points of each branch indicate the health state in which the patient will begin the next 1-month cycle. For each cycle, an evaluation is made whether the patient is due to have a device replacement based on an elective replacement indicator (ERI). Estimated probabilities drive chance events that may occur in each cycle (eg, remain well, device failure that may lead to immediate death or remain undetected, death due to other causes, and death due to device replacement procedure when incurred).

The Markov model represents a simulation of a 2-armed clinical trial: in one arm, the device is immediately replaced and in the other arm, replacement is delayed until the otherwise scheduled replacement date, based on the remaining generator life. In the immediate replacement arm, the patient enters a health state denoted as device replacement, indicating that the device will be replaced. The patient is at risk of procedure-related death, including death due to complications such as systemic infection or lead extraction, as well as risks unrelated to the implanted device. With successful device replacement, the patient enters the state of device functioning. The patient remains in this state until he/she either experiences a random device malfunction unrelated to the advisory or requires another device replacement procedure (based on the estimated generator longevity). In the former situation, the patient may immediately die, may develop symptoms prompting diagnosis of device failure and resulting in device replacement, or may continue follow-up with an undetected device failure (transitioning to the state of undetected device failure). In the undetected device failure state, the patient is faced with a risk of an event-triggered death (eg, such as ventricular fibrillation) due to device failure. Otherwise, the patient continues until generator battery depletion or failure detection, both of which lead to device replacement. For the initial and any subsequent transitions to the device replacement state, the patient undergoes a device replacement procedure with its associated risks; if surviving the procedure, the patient returns to the device functioning state.

In the continued monitoring arm, the patient first enters the device under advisory state, wherein he/she is at risk of the aforementioned events in the immediate replacement arm, with the exception of the risk from initial device replacement procedure. For the remaining duration of the recalled generator's life, the patient faces an additional risk of death due to device failure related to the advisory. This risk is in addition to the underlying random device malfunction risk (ie, risk unrelated to the advisory). We assumed that a patient with a device failure would undergo the same transitions as with a randomly occurring device failure and have identical risks as in the immediate replacement arm. In the event of a non–life-threatening asymptomatic device failure, the patient transitions to the undetected device failure state. This state also considers variations in follow-up frequency in an effort to detect device malfunction prior to symptomatic presentation. Conversely, without device failure, the patient continues in the same state (device under advisory) as long as the generator life of the recalled device is not depleted. When the generator reaches end of life, the patient undergoes device replacement (transitioning to the device replacement state), then moves on to the device functioning state. The latter state is identical to the same state in the immediate replacement arm. In all states, the patient is subject to an age-specific probability of death.

Model Inputs

We obtained estimates and plausible ranges of event probabilities from published articles and expert opinion (Table 2). The baseline monthly probability of death was taken from published life tables for the 2002 US population29 and adjusted for excess cardiovascular mortality associated with being an individual with an implanted cardiovascular device.2528,30 Monthly mortality transition probabilities were calculated using the declining exponential approximation of life expectancy.30

Table Graphic Jump LocationTable 2. Event Probabilities of Health States for Each Scenario

Recently published data were used to estimate the range of failure rates associated with an advisory (reported from 0.009%-2.6%13,9,22,23) as well as the underlying risk of random malfunction (0.1% per year for pacemakers and 2.0% for ICDs2,3). We expanded these ranges in our analysis to cover additional possibilities. We assumed that the probability of device malfunction remained constant throughout the life of the device.

Analysis and Assumptions

For each management option, life expectancy was calculated by tracking a large number of hypothetical individuals whose life history was described by the Markov diagram, summing the outcomes for each hypothetical history, and averaging over the number of modeled individuals. Effectively, patients traverse the model tree cycle by cycle until they die. The average life span is the life expectancy for the cohort. The strategy with the highest value for life expectancy is designated the preferred strategy.

The cycle length for the Markov models is 1 month. The decision analysis is based on this decision being made when presented with a patient with a device under advisory. We assumed that event rates were constant over time. We also assumed that the issue of an active recall or advisory does not affect the underlying random failure rates and that the patient would not be faced with another recalled device beyond the initial device but could experience device failure based on the random device failure rates.

Given the uncertainty in the estimates and to examine a range of possible scenarios, we performed sensitivity analyses6,7 by varying each input parameter over its plausible range (Table 2) while keeping all other parameters constant (1-way sensitivity analysis). Potentially key variables were identified by the existence of a threshold value, a value within the plausible range at which the preferred strategy switches from one to the other. We further evaluated combinations of these variables as well as variables that were of particular clinical interest (multiway sensitivity analysis). In addition, we performed probabilistic sensitivity analysis by examining a large number of runs of the model, for each run assigning a random value to the model inputs based on a beta distribution with the baseline value as the mean and the plausible range as the 95% confidence range. The result is an estimate of the confidence—given the inherent uncertainty in the model inputs—that one strategy is preferred over the other.

All modeling and analyses were carried out on a personal computer using a software package designed for medical decision making applications (TreeAge Pro, version 2005, release 1.3, TreeAge Software Inc, Williamstown, Mass). An external decision analysis expert reviewed the decision model and programming. To further validate the model, we set model inputs at extreme values (beyond the plausible ranges) and compared the output with logical predictions. Examples of the tests and logical predications included the following: (1) in all instances with zero risk of immediate and event-based death due to device failure, regardless of advisory failure rates, continued monitoring is preferred; (2) in all instances with zero risk of advisory device failure, regardless of a risk of death due to device failure, continued monitoring is preferred; and (3) in all instances with zero procedural mortality, with any (even minimal) risk of advisory device failure and a risk of death due to device failure, immediate replacement is preferred.

Results are presented by the 3 classes corresponding to the principle concern should failure occur: (1) immediate death, (2) delayed (event-dependent) death, and (3) symptoms without risk of death.

Failure Associated With Immediate Death

Scenarios 1 and 2 typify patients characterized by a high probability of immediate death with device failure due to pacemaker dependence. For both scenarios, the decision to replace the device is strongly influenced by the advisory failure rate and, to a lesser extent, probability of immediate death given failure and remaining generator life (Figure 2A). As indicated by the upper shaded territory for each scenario, advisory failure rates of less than 3 in 10 000 favor continued monitoring rather than immediate replacement. For these scenarios, replacement becomes the preferred strategy when advisory failure rates are in the range of 3 in 1000 to 3 in 10 000, depending on the procedural mortality rate. For example, at the highest procedural mortality, device replacement is preferred with advisory failure rate estimates greater than 3 in 1000.

Figure 2. Three-Way Threshold Analyses Identifying Combinations of Values for Which Device Replacement or Continued Observation Is the Preferred Strategy
Graphic Jump Location

Variables include estimated advisory device failure rates (shown on the y-axis in logarithmic scale), remaining pulse generator life, device dependency (x-axis), and procedural mortality, with various threshold lines for differing procedural mortality rates. Shaded regions apply to procedural mortality risk of 0.5% (panels A and B) and 0.3% (panel C); for higher and lower procedure mortality, the regions favoring immediate replacement will contract (shift to the higher threshold line) or expand (shift to the lower threshold line), respectively. Panel A evaluates scenarios 1 and 2, failures associated primarily with concern for immediate death, for either pacemakers or implantable cardioverter-defibrillators with pacemaker dependence. Panel B evaluates scenarios 3 and 4, where failure is associated primarily with concern for delayed (event-dependent) death, for patients with implantable cardioverter-defibrillator for primary or secondary prevention. Panel C evaluates scenarios 5 and 6, failures associated primarily with concern for symptoms without a significant risk of death, typically for patients with pacemakers for various stable bradyarrhythmias.

Beyond the risk of immediate death, scenario 2 also has a higher risk of delayed death due to the need for episodic therapy for ventricular tachycardia or ventricular fibrillation. The analysis shows that this additional risk does not have a significant effect on the final decision to replace the ICD. For both scenarios 1 and 2, higher procedural mortality rates increase the advisory failure rate that would be required for device replacement to be preferred. When the risk of immediate death with device failure varies from 40% to 60%, there is only a slight increase in the preference for replacement.

Given the high risk of early death or symptoms with device failure, the decision in this group of patients is insensitive to the follow-up schedule and the underlying random failure rates. Remaining battery life is significant only when this residual is small (<10% of the total generator life), when continued monitoring of devices is preferred until time for elective replacement.

In a probabilistic sensitivity analysis evaluating these scenarios with hypothetical device advisories, similar results were observed (Table 3; scenarios 1 and 2). At higher advisory failure rates (2.5% for ICDs and 0.5% for pacemakers), 100% of the simulations favored device replacement in situations of pacemaker dependence, regardless of variability in the procedure mortality, underlying random failure rates, or probability of immediate death. With lower failure rates (reflecting the scenario approaching the threshold lines in Figure 2), the preference for immediate replacement strategy becomes less certain. Device replacement is preferred in more instances, with a lower advisory rate in pacemakers compared with ICDs (advisory rate of 0.075% in pacemakers and 0.25% in ICDs, yielding decisions to replace the device in 76% and 58% of simulations, respectively). This is attributed to the lower procedure mortality risks with pacemaker generator replacement. The mean prolongation in survival with the preferred strategy ranges from 1 week to 46 weeks.

Table Graphic Jump LocationTable 3. Optimal Management Decisions and Probabilistic Outcomes
Failure Associated With Delayed (Event-Dependent) Death

Scenarios 3 and 4 typify patients who are dependent on their device to provide therapy for ventricular tachyarrhythmias (ie, risk of delayed death due to device failure). Both scenarios are characterized by a relatively low probability of immediate death with device failure and a low likelihood of symptoms developing with device failure. As such, these scenarios also have the potential for undetected failure because device therapy is not required continuously.

Results of our analysis (Figure 2B) reveal that advisory failure rates and risk of delayed death with device failure directly affect the decision. Because of the relatively low risk of death with device failure due to advisory, failure rates greater than 3 in 100 are required for elective device replacement to be preferred. As procedure mortality rates decrease to 0.1% or risk of fatal arrhythmias increase to near 20% per year, there is a decrease in the estimated advisory rate to greater than 1 in 100 needed for replacement to be preferred. Similar to the first group, lower underlying random failure rates and higher generator life decrease the advisory failure rate favoring replacement. A more frequent follow-up schedule appears to provide a slight relative advantage to continue monitoring devices but does not significantly alter the preferred decision.

The probabilistic analysis results support the multiway sensitivity analysis (Table 3; scenarios 3 and 4). At a lower advisory failure rate of 0.25%, in both scenarios continued monitoring of devices is favored in the majority of simulations. Conversely, an advisory with a higher failure rate of 2.5% when an ICD is implanted for secondary prevention (scenario 4), immediate replacement is favored in the majority of simulations. In scenario 3, involving an advisory with a 2.5% failure rate for an ICD implanted for primary prevention, the decision is effectively a toss-up, with about half of simulations favoring device replacement. The mean prolongation of survival with the preferred strategy ranges from 1 week to 8 weeks.

Failure Associated With Symptoms Without Significant Risk of Death

Scenarios 5 and 6 typify patients whose pacemakers have been implanted primarily because of a transient bradyarrhythmia leading to cardiac symptoms. Typically, these patients have low risks of immediate and delayed death due to device failure. In general, device failure in these patients will result in a large proportion of patients developing symptoms that alert them and their physicians to possible device malfunction. It is important to note that there are limited data available for this group in terms of risk of death due to device failure, so model inputs were based on expert estimates.

Analysis of model results (Figure 2C) showed that in a wide range of conditions, devices in this group do not require replacement except at advisory failure rates of 1 in 25 or higher. Only at high failure rates are other factors, such as procedural mortality and probability of delayed death, important considerations. Of note, a remaining generator life of less than 2 years strongly influences the decision to continue monitoring most devices.

For this last group, the probabilistic analysis consistently supports continued monitoring for both scenarios (Table 3; scenarios 5 and 6). The simulations show that this preference is relatively certain, with more than 90% of simulations favoring continued monitoring. The mean prolongation in survival with the preferred strategy ranges from 1 week to 2 weeks.

Additional Sensitivity Analyses

A total of fifty-six 1-way sensitivity analyses were performed for all combinations of input variables among the 6 scenarios (available at http://www.clinpol.mc.duke.edu). In the sensitivity analysis, 2 variables exhibited a differential effect on the life expectancy associated with the strategies. These variables were the remaining generator life and the advisory failure rates. In particular, the analysis for advisory failure rates revealed threshold values that increased among scenarios as the untoward effects of failure becomes less serious (failures per year of 0.08%, 0.16%, 7.9%, 4.0%, 4.0%, and 7.9% for scenarios 1-6, respectively). Continued monitoring is the preferred strategy with advisory failure rates less than the threshold value, whereas immediate replacement is preferred for rates higher than the threshold values.

Additional sensitivity analyses were used to further evaluate the impact of advisory failure rates, specifically in situations of increasing rates over time. An extreme analysis with rates doubling on a yearly basis found that there was nearly a 10-fold decrease in threshold values compared with the constant base case (data available at http://www.clinpol.mc.duke.edu). However, given limited clinical data on variable or increasing advisory failure rates, subsequent analyses used constant rates. Also, several variables of clinical importance did not exhibit a major impact on the relative strategy preference in the 1-way analysis but were included in the multiway analysis because of their clinical relevance.

The major conclusions of our analysis are that (1) the decision to replace or follow up on a device under advisory is most influenced by the likely impact of device failure, the estimated advisory failure rate, and, to a lesser degree, the procedural mortality rate; (2) the remaining generator life and the patient's age and sex have minimal influence on the decision to replace a device; (3) device replacement appears to be preferred at lower advisory failure rates (approximately 3/10 000) in pacemaker-dependent situations; and (4) in all other instances, higher advisory failure rates are needed to warrant replacement.

Device recalls and advisories confront clinicians with competing risks without a clear management approach described in the medical literature. The notion that replacement or removal of devices may result in a higher mortality than careful follow-up was realized in the management of Accufix pacemaker leads (Telectronics Pacing, Englewood, Colo) as well as in a recently reported Canadian experience.9,31 In the case of the Accufix pacemaker leads, a conservative management strategy was associated with lower mortality than lead extraction in most patients.31 More recently, Gould et al9 reported a high incidence of serious complications, including death, associated with elective device replacements prompted by advisories.

Our analysis shows that device replacement in the setting of an advisory is not inconsequential and frequently has a greater risk than continued device follow-up. The decision to replace a recalled device should be based primarily on the device failure rate, the degree of patient dependency, and the anticipated procedural mortality from device replacement. Recognizing that the decision to replace a device will be made on a case-by-case basis, this analysis provides a framework for rational discussions with and risk stratification for individual patients.

In this analysis, we considered the potential factors in deciding whether to immediately replace a device under advisory or to continue without replacement. An important issue that we considered is that once the advisory device is replaced, there continues to be a risk of future random failure with the new device. If random failure rates are relatively high, then the impetus to replace a device under advisory is decreased. As random failure rates decrease, the decision to replace a device is altered.

Our analysis has several notable clinical implications. First, the clinical situations favoring device replacement are likely more limited than most physicians estimate, and indiscriminant device replacement may result in more deaths than lives saved. Second, physician reporting is paramount to providing accurate estimates of device failure rates. This assumes that the US Food and Drug Administration, physicians, and manufacturers provide better patient ascertainment data, timely reporting, and more accurate estimates of device failure rates. A stringent, mandatory, simplified reporting system may be needed to accomplish this important goal. Third, individual centers must obtain and report accurate procedural complication rates to determine the real risk of device replacement. Finally, our analysis shows that decisions about removing pacemakers and defibrillators are based on similar factors, the difference being the numbers used in the analysis for degree of device dependency, risk of infection, and risk of lead removal.

Another factor accounted for in our model is the impact of increased patient follow-up. Although our data do not show a clear impact, increased follow-up does alter the failure rate needed for device replacement if there is a high incidence of device failure without immediate death or with delayed death. It is likely that future new technologies of remote follow-up will have an even larger impact favoring device monitoring when device failure is unlikely to result in immediate death.

There are several limitations to our analysis. The model is derived from published data not specifically related to device recalls. The ranges of event rates may not cover all possibilities, and there will be some centers that may have higher or lower risks of infection and death due to lead extraction. Additionally, there is no data on mortality from device failure in patients with different types of electrical heart disease. Our model did not estimate the psychological burden resulting from the knowledge that a device may fail. This is likely to be a factor in many cases, and more studies are needed to appropriately quantify this impact.32 Additionally, we assumed a constant device malfunction rate over time, acknowledging that for certain devices the probability of device malfunction may increase with the age of the device and that this could significantly affect the decision.22 Life expectancy was based on life-table data adjusted for excess mortality associated with cardiovascular disease. We examined a broad range of estimates of excess mortality and found that this did not substantially alter our results. Notably, one benefit of modeling is that a model provides an opportunity to improve estimates of the likely benefit of replacement vs continued observation as estimates of the various risks improve.

In conclusion, as the number of implantable devices increase, device failures and advisories will remain a part of routine practice. Having a rational mechanism to approach these patients is critical to patient care. The risks of replacing devices are not insignificant and outweigh all but very high risks of death due to device malfunction. Ultimately, this decision should be informed by the estimated device malfunction rate, anticipated consequences of device failure, and an individual center's procedural risk of complications from generator change, in conjunction with patient preferences and tolerance of risk.

Corresponding Author: Kenneth A. Ellenbogen, MD, Medical College of Virginia, PO Box 980053, Richmond, VA 23298-0053 (kellenbogen@pol.net or kaellenb@vcu.edu).

Author Contributions: Drs Amin and Matchar had full access to all of the data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis.

Study concept and design: Amin, Matchar, Wood, Ellenbogen.

Acquisition of data: Amin, Ellenbogen.

Analysis and interpretation of data: Amin, Matchar, Ellenbogen.

Drafting of the manuscript: Amin, Matchar, Wood, Ellenbogen.

Critical revision of the manuscript for important intellectual content: Matchar, Ellenbogen.

Statistical analysis: Amin, Matchar.

Administrative, technical, or material support: Matchar, Ellenbogen.

Study supervision: Matchar, Wood.

Financial Disclosures: Dr Matchar is the director of the Duke Evidence-based Practice Center, commissioned to work with Center for Medicare and Medical Services for purposes of technology assessment. Drs Wood and Ellenbogen participate in device-related clinical research supported by grants from Medtronic, Guidant, St Jude, and Cameron Health. Dr Ellenbogen reports having received honoraria from Medtronic, Guidant, St Jude, and Sorin Biomedica. No other disclosures were reported. None of the outside companies that are mentioned were in any way involved with this work.

Acknowledgment: We thank Shalani Kulasingam, PhD, MPH, from the Center for Clinical Health Policy Research, Duke University, for reviewing the decision analysis model, and Emily Matchar for assistance with copy editing of an early version of the manuscript.

Maisel WH. Safety issues involving medical devices: implications of recent implantable cardioverter-defibrillator malfunctions.  JAMA. 2005;294:955-958
PubMed   |  Link to Article
Maisel WH, Moynahan M, Zuckerman BD.  et al.  Pacemaker and ICD generator malfunctions: analysis of Food and Drug Administration annual reports.  JAMA. 2006;295:1901-1906
PubMed   |  Link to Article
Maisel WH. Pacemaker and ICD generator reliability: meta-analysis of device registries.  JAMA. 2006;295:1929-1934
PubMed   |  Link to Article
Auricchio A, Gropp M, Ludgate S.  et al.  European Heart Rhythm Association guidance document on cardiac rhythm management product performance.  Europace. 2006;8:313-322
PubMed   |  Link to Article
Maisel WH. Physician management of pacemaker and implantable cardioverter-defibrillator advisories.  Pacing Clin Electrophysiol. 2004;27:437-442
PubMed   |  Link to Article
Weinstein MC, Fineberg HV, Elstein AS.  et al.  Clinical Decision Analysis. Philadelphia, Pa: WB Saunders; 1980:131-166
Sonnenberg FA, Beck JR. Markov models in medical decision making: a practical guide.  Med Decis Making. 1993;13:322-338
PubMed   |  Link to Article
Wild DM, Fisher JD, Kim SG, Ferrick KJ, Gross JN, Palma EC. Pacemakers and implantable cardioverter defibrillators: device longevity is more important than smaller size: the patient's viewpoint.  Pacing Clin Electrophysiol. 2004;27:1526-1529
PubMed   |  Link to Article
Gould PA, Krahn AD.Canadian Heart Rhythm Society Working Group on Device Advisories.  Complications associated with implantable cardioverter-defibrillator replacement in response to device advisories.  JAMA. 2006;295:1907-1911
PubMed   |  Link to Article
Thomas AC, Moser SA, Smutka ML, Wilson PA. Implantable defibrillation: eight years clinical experience.  Pacing Clin Electrophysiol. 1988;11:2053-2058
PubMed   |  Link to Article
Smith PN, Vidaillet HJ, Hayes JJ.  et al.  Infection with nonthoracotomy implantable cardioverter defibrillators: can these be prevented?  Pacing Clin Electrophysiol. 1998;21:42-55
PubMed   |  Link to Article
Mela T, McGovern BA, Garan H.  et al.  Long-term infection rates associated with the pectoral versus abdominal approach to cardioverter defibrillator implants.  Am J Cardiol. 2001;88:750-753
PubMed   |  Link to Article
Gold MR, Peters RW, Johnson JW, Shorofsky SR. Complications associated with pectoral cardioverter-defibrillator implantation: comparison of subcutaneous and submuscular approaches.  J Am Coll Cardiol. 1996;28:1278-1282
PubMed   |  Link to Article
Alter P, Waldhans S, Plachta E, Moosdorf R, Grimm W. Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients.  Pacing Clin Electrophysiol. 2005;28:926-932
PubMed   |  Link to Article
Wiegand UK, LeJeune D, Boguschewski F.  et al.  Pocket hematoma after pacemaker or implantable cardioverter defibrillator surgery: influence of patient morbidity, operation strategy, and perioperative antiplatelet/anticoagulation therapy.  Chest. 2004;126:1177-1186
PubMed   |  Link to Article
Zaim S, Connolly M, Roman-Gonzalez J, Walter P, Craver J, Jones E. Perioperative complications of cardioverter-defibrillator implantation: the Emory experience.  Am J Med Sci. 1994;307:185-189
PubMed   |  Link to Article
Harcombe AA, Newell SA, Ludman PF.  et al.  Late complications following permanent pacemaker implantation or elective unit replacement.  Heart. 1998;80:240-244
PubMed
Chauhan A, Grace AA, Newell SA.  et al.  Early complications after dual chamber versus single chamber pacemaker implantation.  Pacing Clin Electrophysiol. 1994;17:2012-2015
PubMed   |  Link to Article
Ellenbogen KA, Hellkamp AS, Wilkoff BL.  et al.  Complications arising after implantation of DDD pacemakers: the MOST experience.  Am J Cardiol. 2003;92:740-741
PubMed   |  Link to Article
Aggarwal RK, Connelly DT, Ray SG, Ball J, Charles RG. Early complications of permanent pacemaker implantation: no difference between dual and single chamber systems.  Br Heart J. 1995;73:571-575
PubMed   |  Link to Article
Tobin K, Stewart J, Westveer D, Frumin H. Acute complications of permanent pacemaker implantation: their financial implication and relation to volume and operator experience.  Am J Cardiol. 2000;85:774-776, A9
PubMed   |  Link to Article
Medtronic.  Marquis Patient Management InformationMinneapolis, Minn: Medtronic; 2005
 Guidant Advisory Update: Contak Renewal & Contak Renewal 2, Models H135 and H155. Indianapolis, Ind: Guidant; 2005
Mirowski M, Reid PR, Mower MM.  et al.  Termination of malignant ventricular arrhythmias with an implantable automatic defibrillator in human beings.  N Engl J Med. 1980;303:322-324
PubMed   |  Link to Article
Antiarrhythmics Versus Implantable Defibrillators (AVID) Investigators.  A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias.  N Engl J Med. 1997;337:1576-1583
PubMed   |  Link to Article
Bristow MR, Saxon LA, Boehmer J.  et al. Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) Investigators.  Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure.  N Engl J Med. 2004;350:2140-2150
PubMed   |  Link to Article
Moss AJ, Zareba W, Hall WJ.  et al. Multicenter Automatic Defibrillator Implantation Trial II Investigators.  Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction.  N Engl J Med. 2002;346:877-883
PubMed   |  Link to Article
Bardy GH, Lee KL, Mark DB.  et al. Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) Investigators.  Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure.  N Engl J Med. 2005;352:225-237
PubMed   |  Link to Article
Kochanek KD, Murphy SL, Anderson RN, Scott C. Deaths: final data for 2002.  Natl Vital Stat Rep. 2004;53:1-115
PubMed
Beck JR, Pauker SG, Gottlieb JE, Klein K, Kassirer JP. A convenient approximation of life expectancy (the “DEALE”), II: use in medical decision making.  Am J Med. 1982;73:889-897
PubMed   |  Link to Article
Kay GN, Brinker JA, Kawanishi DT.  et al.  Risks of spontaneous injury and extraction of an active fixation pacemaker lead: report of the Accufix Multicenter Clinical Study and Worldwide Registry.  Circulation. 1999;100:2344-2352
PubMed   |  Link to Article
Undavia M, Goldstein NE, Cohen P.  et al.  Impact of implantable cardioverter defibrillator recall on patients' anxiety, depression, and quality of life.  Heart Rhythm. 2006;3:S177-S178
Link to Article

Figures

Figure 1. Simplified Diagram of the Markov Decision Analysis Model
Graphic Jump Location

Squares indicate decision nodes, choices facing the decision maker. Circles indicate chance nodes, events that have multiple possible outcomes and are not in the decision maker’s control. The square at the left indicates 2 initial management options, immediate device replacement and continued monitoring. After the initial decision, each option enters a Markov process, which leads to 1 of many health states. The initial path within the Markov process is forced (bold), device replacement state for the immediate replacement decision and device under advisory state for the continued monitoring decision. The Markov cycle represents a single 1-month cycle of the model (ie, 1 month of a patient's life), and the end points of each branch indicate the health state in which the patient will begin the next 1-month cycle. For each cycle, an evaluation is made whether the patient is due to have a device replacement based on an elective replacement indicator (ERI). Estimated probabilities drive chance events that may occur in each cycle (eg, remain well, device failure that may lead to immediate death or remain undetected, death due to other causes, and death due to device replacement procedure when incurred).

Figure 2. Three-Way Threshold Analyses Identifying Combinations of Values for Which Device Replacement or Continued Observation Is the Preferred Strategy
Graphic Jump Location

Variables include estimated advisory device failure rates (shown on the y-axis in logarithmic scale), remaining pulse generator life, device dependency (x-axis), and procedural mortality, with various threshold lines for differing procedural mortality rates. Shaded regions apply to procedural mortality risk of 0.5% (panels A and B) and 0.3% (panel C); for higher and lower procedure mortality, the regions favoring immediate replacement will contract (shift to the higher threshold line) or expand (shift to the lower threshold line), respectively. Panel A evaluates scenarios 1 and 2, failures associated primarily with concern for immediate death, for either pacemakers or implantable cardioverter-defibrillators with pacemaker dependence. Panel B evaluates scenarios 3 and 4, where failure is associated primarily with concern for delayed (event-dependent) death, for patients with implantable cardioverter-defibrillator for primary or secondary prevention. Panel C evaluates scenarios 5 and 6, failures associated primarily with concern for symptoms without a significant risk of death, typically for patients with pacemakers for various stable bradyarrhythmias.

Tables

Table Graphic Jump LocationTable 1. Scenarios of Device Failure Characteristics and Risk of Events
Table Graphic Jump LocationTable 2. Event Probabilities of Health States for Each Scenario
Table Graphic Jump LocationTable 3. Optimal Management Decisions and Probabilistic Outcomes

References

Maisel WH. Safety issues involving medical devices: implications of recent implantable cardioverter-defibrillator malfunctions.  JAMA. 2005;294:955-958
PubMed   |  Link to Article
Maisel WH, Moynahan M, Zuckerman BD.  et al.  Pacemaker and ICD generator malfunctions: analysis of Food and Drug Administration annual reports.  JAMA. 2006;295:1901-1906
PubMed   |  Link to Article
Maisel WH. Pacemaker and ICD generator reliability: meta-analysis of device registries.  JAMA. 2006;295:1929-1934
PubMed   |  Link to Article
Auricchio A, Gropp M, Ludgate S.  et al.  European Heart Rhythm Association guidance document on cardiac rhythm management product performance.  Europace. 2006;8:313-322
PubMed   |  Link to Article
Maisel WH. Physician management of pacemaker and implantable cardioverter-defibrillator advisories.  Pacing Clin Electrophysiol. 2004;27:437-442
PubMed   |  Link to Article
Weinstein MC, Fineberg HV, Elstein AS.  et al.  Clinical Decision Analysis. Philadelphia, Pa: WB Saunders; 1980:131-166
Sonnenberg FA, Beck JR. Markov models in medical decision making: a practical guide.  Med Decis Making. 1993;13:322-338
PubMed   |  Link to Article
Wild DM, Fisher JD, Kim SG, Ferrick KJ, Gross JN, Palma EC. Pacemakers and implantable cardioverter defibrillators: device longevity is more important than smaller size: the patient's viewpoint.  Pacing Clin Electrophysiol. 2004;27:1526-1529
PubMed   |  Link to Article
Gould PA, Krahn AD.Canadian Heart Rhythm Society Working Group on Device Advisories.  Complications associated with implantable cardioverter-defibrillator replacement in response to device advisories.  JAMA. 2006;295:1907-1911
PubMed   |  Link to Article
Thomas AC, Moser SA, Smutka ML, Wilson PA. Implantable defibrillation: eight years clinical experience.  Pacing Clin Electrophysiol. 1988;11:2053-2058
PubMed   |  Link to Article
Smith PN, Vidaillet HJ, Hayes JJ.  et al.  Infection with nonthoracotomy implantable cardioverter defibrillators: can these be prevented?  Pacing Clin Electrophysiol. 1998;21:42-55
PubMed   |  Link to Article
Mela T, McGovern BA, Garan H.  et al.  Long-term infection rates associated with the pectoral versus abdominal approach to cardioverter defibrillator implants.  Am J Cardiol. 2001;88:750-753
PubMed   |  Link to Article
Gold MR, Peters RW, Johnson JW, Shorofsky SR. Complications associated with pectoral cardioverter-defibrillator implantation: comparison of subcutaneous and submuscular approaches.  J Am Coll Cardiol. 1996;28:1278-1282
PubMed   |  Link to Article
Alter P, Waldhans S, Plachta E, Moosdorf R, Grimm W. Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients.  Pacing Clin Electrophysiol. 2005;28:926-932
PubMed   |  Link to Article
Wiegand UK, LeJeune D, Boguschewski F.  et al.  Pocket hematoma after pacemaker or implantable cardioverter defibrillator surgery: influence of patient morbidity, operation strategy, and perioperative antiplatelet/anticoagulation therapy.  Chest. 2004;126:1177-1186
PubMed   |  Link to Article
Zaim S, Connolly M, Roman-Gonzalez J, Walter P, Craver J, Jones E. Perioperative complications of cardioverter-defibrillator implantation: the Emory experience.  Am J Med Sci. 1994;307:185-189
PubMed   |  Link to Article
Harcombe AA, Newell SA, Ludman PF.  et al.  Late complications following permanent pacemaker implantation or elective unit replacement.  Heart. 1998;80:240-244
PubMed
Chauhan A, Grace AA, Newell SA.  et al.  Early complications after dual chamber versus single chamber pacemaker implantation.  Pacing Clin Electrophysiol. 1994;17:2012-2015
PubMed   |  Link to Article
Ellenbogen KA, Hellkamp AS, Wilkoff BL.  et al.  Complications arising after implantation of DDD pacemakers: the MOST experience.  Am J Cardiol. 2003;92:740-741
PubMed   |  Link to Article
Aggarwal RK, Connelly DT, Ray SG, Ball J, Charles RG. Early complications of permanent pacemaker implantation: no difference between dual and single chamber systems.  Br Heart J. 1995;73:571-575
PubMed   |  Link to Article
Tobin K, Stewart J, Westveer D, Frumin H. Acute complications of permanent pacemaker implantation: their financial implication and relation to volume and operator experience.  Am J Cardiol. 2000;85:774-776, A9
PubMed   |  Link to Article
Medtronic.  Marquis Patient Management InformationMinneapolis, Minn: Medtronic; 2005
 Guidant Advisory Update: Contak Renewal & Contak Renewal 2, Models H135 and H155. Indianapolis, Ind: Guidant; 2005
Mirowski M, Reid PR, Mower MM.  et al.  Termination of malignant ventricular arrhythmias with an implantable automatic defibrillator in human beings.  N Engl J Med. 1980;303:322-324
PubMed   |  Link to Article
Antiarrhythmics Versus Implantable Defibrillators (AVID) Investigators.  A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias.  N Engl J Med. 1997;337:1576-1583
PubMed   |  Link to Article
Bristow MR, Saxon LA, Boehmer J.  et al. Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) Investigators.  Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure.  N Engl J Med. 2004;350:2140-2150
PubMed   |  Link to Article
Moss AJ, Zareba W, Hall WJ.  et al. Multicenter Automatic Defibrillator Implantation Trial II Investigators.  Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction.  N Engl J Med. 2002;346:877-883
PubMed   |  Link to Article
Bardy GH, Lee KL, Mark DB.  et al. Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) Investigators.  Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure.  N Engl J Med. 2005;352:225-237
PubMed   |  Link to Article
Kochanek KD, Murphy SL, Anderson RN, Scott C. Deaths: final data for 2002.  Natl Vital Stat Rep. 2004;53:1-115
PubMed
Beck JR, Pauker SG, Gottlieb JE, Klein K, Kassirer JP. A convenient approximation of life expectancy (the “DEALE”), II: use in medical decision making.  Am J Med. 1982;73:889-897
PubMed   |  Link to Article
Kay GN, Brinker JA, Kawanishi DT.  et al.  Risks of spontaneous injury and extraction of an active fixation pacemaker lead: report of the Accufix Multicenter Clinical Study and Worldwide Registry.  Circulation. 1999;100:2344-2352
PubMed   |  Link to Article
Undavia M, Goldstein NE, Cohen P.  et al.  Impact of implantable cardioverter defibrillator recall on patients' anxiety, depression, and quality of life.  Heart Rhythm. 2006;3:S177-S178
Link to Article

Letters

CME
Meets CME requirements for:
Browse CME for all U.S. States
Accreditation Information
The American Medical Association is accredited by the Accreditation Council for Continuing Medical Education to provide continuing medical education for physicians. The AMA designates this journal-based CME activity for a maximum of 1 AMA PRA Category 1 CreditTM per course. Physicians should claim only the credit commensurate with the extent of their participation in the activity. Physicians who complete the CME course and score at least 80% correct on the quiz are eligible for AMA PRA Category 1 CreditTM.
Note: You must get at least of the answers correct to pass this quiz.
You have not filled in all the answers to complete this quiz
The following questions were not answered:
Sorry, you have unsuccessfully completed this CME quiz with a score of
The following questions were not answered correctly:
Commitment to Change (optional):
Indicate what change(s) you will implement in your practice, if any, based on this CME course.
Your quiz results:
The filled radio buttons indicate your responses. The preferred responses are highlighted
For CME Course: A Proposed Model for Initial Assessment and Management of Acute Heart Failure Syndromes
Indicate what changes(s) you will implement in your practice, if any, based on this CME course.

Multimedia

Some tools below are only available to our subscribers or users with an online account.

Web of Science® Times Cited: 31

Related Content

Customize your page view by dragging & repositioning the boxes below.

Articles Related By Topic
Related Collections
PubMed Articles
JAMAevidence.com

Users' Guides to the Medical Literature
Table 9.2-3 Refuted Evidence From Observational Studiesa