From the Centers for Disease Control and Prevention |

Pseudomonas aeruginosa FREE

JAMA. 2006;296(19):2308-2309. doi:10.1001/jama.296.19.2308.
Text Size: A A A
Published online

PSEUDOMONAS AERUGINOSA INFECTIONS ASSOCIATED WITH TRANSRECTAL ULTRASOUND-GUIDED PROSTATE BIOPSIES—GEORGIA, 2005

MMWR. 2006;55:776-777

Transrectal ultrasound (TRUS)-guided prostate biopsies are among the most common outpatient diagnostic procedures performed in urology clinics, with an estimated 624,000 performed annually in the United States (CDC, unpublished data, 2006). The procedures generally are performed in follow-up to elevated levels of prostate-specific antigen or abnormal digital rectal examinations.1 Septicemia has been reported as a rare complication of the procedure.2 This report summarizes an investigation of four cases of Pseudomonas aeruginosa infection after TRUS-guided prostate biopsies in which contamination of the equipment was the likely source. The findings underscore the need to adhere to recommendations for the cleaning and disinfection of TRUS-guided prostate biopsy equipment.

On July 28, 2005, a urologist notified the Georgia Department of Human Resources, Division of Public Health (GDPH) regarding four patients who were hospitalized with P. aeruginosa infections within 6 days of outpatient TRUS-guided prostate biopsies performed at a clinic. All procedures were halted at the clinic pending the investigation. The four patients were white, non-Hispanic men aged 57-71 years who had undergone the biopsy procedure during July 20-26, 2005. They were the only patients who had TRUS-guided prostate biopsies at the clinic during that period. Subsequently, all four experienced fever and chills and were admitted to the hospital 1-6 days (mean: 2.5 days) after their procedures. Three patients were admitted with diagnosed septicemia and the fourth with a diagnosis of infection. P. aeruginosa was recovered from cultures of blood (one patient), urine (two patients), or blood and urine specimens (one patient). The patients were treated successfully with a combination of intravenous and oral antimicrobial agents during hospitalizations of 2-12 days (mean: 5.8 days).

All procedures had been performed in the clinic by the same urologist and staff members using the following technique. Immediately before each procedure, a new finger cot was fitted over the distal tip of the ultrasound probe, filled with gel to eliminate air bubbles, and secured with an O-ring. A standard condom was then fitted over the finger cot and ultrasound probe and filled with lubricant. Next, a steel, nondisposable needle guide was fitted over the ultrasound probe, finger cot, and first condom. A second condom was fitted over these items and filled with lubricant. Once the ultrasound probe was inserted into the rectum and positioned correctly, the urologist used a spring-loaded biopsy gun to fire a sterile biopsy needle through the needle guide into the prostate, piercing the second condom, to obtain a core of tissue for pathologic analysis. The same needle was withdrawn and reinserted through the needle guide approximately eight times to obtain the needed tissue cores from each patient.

The clinic's standard practice for perioperative prophylaxis included administration of 500 mg of levofloxacin orally the night before the procedure, an enema per rectum 1 hour before the procedure, and 80 mg of gentamicin intramuscularly upon arrival at the clinic on the day of the biopsy. After the procedure, patients were instructed to take 500 mg of levofloxacin orally daily for 3 days.

After each procedure, the ultrasound probe was disinfected by wiping it with a 3.2% glutaraldehyde solution. A syringe was used to flush the steel needle guide first with soap, then with tap water, and, finally, with orthophthaldehyde (OPA), a high-level disinfectant. The needle guide was then soaked in the OPA for a minimum of 15 minutes and usually overnight. Before use, the needle guide was removed from the OPA and rinsed with tap water. A review of the manufacturer's written instructions revealed that the recommended reprocessing method for the needle guide called for first cleaning biologic material from the guide and then sterilizing the guide.

A total of 16 environmental samples were obtained from surfaces, supplies, equipment, and tap water in the clinic during August 5-10, 2005. One grew P. aeruginosa; this was a sample obtained from the narrow lumen of the needle guide after it was removed from OPA disinfectant. This specimen was obtained by scraping the needle guide lumen with a sterile needle and then using the needle to inoculate a sterile swab. All four patient isolates and the isolate obtained from the needle guide had similar antimicrobial susceptibility patterns and were resistant to gentamicin and levofloxacin, the agents used for perioperative prophylaxis. The needle-guide isolate and the three available patient isolates were indistinguishable by pulsed-field gel electrophoresis.

Reported by:

J Gillespie, MPH, KE Arnold, MD, Georgia Dept of Human Resources, Div of Public Health. MA Kainer, MBBS, Tennessee Dept of Health. J Noble-Wang, PhD, B Jensen, MMSc, M Arduino, PhD, J Hageman, MPH, A Srinivasan, MD, National Center for Preparedness, Detection, and Control of Infectious Diseases (proposed), CDC.

CDC Editorial Note:

This report de scribes an investigation of P. aeruginosa infections that were likely related to contamination of TRUS prostate biopsy equipment that had not been adequately cleaned (i.e., by brushing) or properly sterilized and had been rinsed improperly with tap water after reprocessing. The association between the equipment and the infections was indicated by matching the strain of P. aeruginosa from the lumen of the reprocessed needle guide with those strains recovered from the three available patient isolates.

Although infectious complications of TRUS-guided prostate biopsies have been reported,2 contamination of the needle guide has not been previously implicated as the cause of infection. According to the Spaulding system for reprocessing medical devices,3 prostate biopsy needle guides are “critical devices” because the needles that pass through them penetrate sterile tissue. After adequate manual cleaning, critical devices must be sterilized before reuse. Steam sterilization is the preferred method for reprocessing heat-stable medical devices, including many prostate biopsy needle guides. The manufacturers of these guides provide recommendations for sterilization methods that are compatible with the specific devices, and users should review and follow these recommendations.

Manual cleaning to remove biologic material is a necessary first step in reprocessing any medical device; disinfection and sterilization protocols do not work effectively on visibly soiled surfaces. Because the lumens of needle guides and needle-guide support channels and assemblies are long and narrow, manual cleaning is difficult without the use of special equipment designed to clean the device. Manufacturers of reusable prostate needle guides recommend the use of special brushes to clean guides and support channels and assemblies. These brushes must be purchased separately from the needle guides, and a new brush should be used each time the guide is cleaned.

Another recent investigation demonstrates that the failure to properly clean the lumen of a prostate needle guide has not been limited to the cases described in this report. In April 2006, the Veterans Health Administration issued a Patient Safety Alert to all U.S. Department of Veterans Affairs (VA) hospitals stating that a routine environmental inspection at a urology clinic revealed that the lumen of a needle guide of a reusable, reprocessed, TRUS transducer assembly was soiled.* The ensuing investigation determined that brushes were not being used to clean the lumen of the needle guide. All VA hospitals were instructed to review procedures for reprocessing this equipment, and other VA facilities also reported that brushes were not being used. The VA alert has prompted reviews by non-VA health-care systems. In Tennessee, facilities contacted the state health department to report that brushes were not being used to reprocess prostate biopsy needle guides. In response, the Tennessee Department of Health disseminated recommendations from the Food and Drug Administration (FDA) on reprocessing TRUS equipment to hospitals, surgical centers, and urologists.

In the cases described in this report, the practice of rinsing the needle guide in tap water after reprocessing might have contributed to its contamination. P. aeruginosa is well known to colonize tap water and has the ability to form biofilms on medical devices that are difficult to remove. Because tap water is not sterile, it should never be used to rinse medical equipment after reprocessing.

In June 2006, in response to the recent reports of problems with reprocessing prostate biopsy needle guides, FDA issued a Public Health Notification. This notification contains a summary of the recommendations for the proper reprocessing of reusable prostate biopsy equipment.† Health-care providers and their staffs should adhere to both the FDA recommendations and the equipment manufacturer's cleaning instructions.

REFERENCES
Wareing M. Transrectal ultrasound and prostate biopsy clinic.  Nurs Stand. 2004;18:33-37
PubMed
Crundwell MC, Cooke PW, Wallace DM. Patients' tolerance of transrectal ultrasound-guided prostatic biopsy: an audit of 104 cases.  BJU Int. 1999;83:792-795
PubMed   |  Link to Article
Spaulding EH. Chemical disinfection of medical and surgical materials [Chapter 32]. In: Lawrence CA, Block SS, eds. Disinfection, sterilization and preservation. Philadelphia, PA: Lea & Febiger; 1968:517-531

Figures

Tables

References

Wareing M. Transrectal ultrasound and prostate biopsy clinic.  Nurs Stand. 2004;18:33-37
PubMed
Crundwell MC, Cooke PW, Wallace DM. Patients' tolerance of transrectal ultrasound-guided prostatic biopsy: an audit of 104 cases.  BJU Int. 1999;83:792-795
PubMed   |  Link to Article
Spaulding EH. Chemical disinfection of medical and surgical materials [Chapter 32]. In: Lawrence CA, Block SS, eds. Disinfection, sterilization and preservation. Philadelphia, PA: Lea & Febiger; 1968:517-531
CME
Also Meets CME requirements for:
Browse CME for all U.S. States
Accreditation Information
The American Medical Association is accredited by the Accreditation Council for Continuing Medical Education to provide continuing medical education for physicians. The AMA designates this journal-based CME activity for a maximum of 1 AMA PRA Category 1 CreditTM per course. Physicians should claim only the credit commensurate with the extent of their participation in the activity. Physicians who complete the CME course and score at least 80% correct on the quiz are eligible for AMA PRA Category 1 CreditTM.
Note: You must get at least of the answers correct to pass this quiz.
Your answers have been saved for later.
You have not filled in all the answers to complete this quiz
The following questions were not answered:
Sorry, you have unsuccessfully completed this CME quiz with a score of
The following questions were not answered correctly:
Commitment to Change (optional):
Indicate what change(s) you will implement in your practice, if any, based on this CME course.
Your quiz results:
The filled radio buttons indicate your responses. The preferred responses are highlighted
For CME Course: A Proposed Model for Initial Assessment and Management of Acute Heart Failure Syndromes
Indicate what changes(s) you will implement in your practice, if any, based on this CME course.

Multimedia

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Articles Related By Topic
Related Collections
PubMed Articles
JAMAevidence.com

The Rational Clinical Examination
Evidence Summary and Review 2

The Rational Clinical Examination
Location of Paracentesis and Ultrasound Guidance