Original Contribution |

Sirolimus- vs Paclitaxel-Eluting Stents in De Novo Coronary Artery Lesions:  The REALITY Trial: A Randomized Controlled Trial FREE

Marie-Claude Morice, MD; Antonio Colombo, MD; Bernhard Meier, MD; Patrick Serruys, MD, PhD; Corrado Tamburino, MD; Giulio Guagliumi, MD; Eduardo Sousa, MD, PhD; Hans-Peter Stoll, MD; for the REALITY Trial Investigators
[+] Author Affiliations

Author Affiliations: Institut Cardiovasculaire Paris Sud, Massy, France (Dr Morice); Centro Cuore Colombus and San Raffaele Hospital, Milan, Italy (Dr Colombo); University Hospital, Bern, Switzerland (Dr Meier); Erasmus Medical Center, Rotterdam, the Netherlands (Dr Serruys); Ferrarotto Hospital, Catania, Italy (Dr Tamburino); Azienda Ospedaliera Ospedali Riuniti di Bergamo, Italy (Dr Guagliumi); Instituto Dante Pazzanese, Sao Paulo, Brazil (Dr Sousa); and Cordis Clinical Research Europe, Waterloo, Belgium (Dr Stoll).

More Author Information
JAMA. 2006;295(8):895-904. doi:10.1001/jama.295.8.895.
Text Size: A A A
Published online

Context Compared with bare metal stents, sirolimus-eluting and paclitaxel-eluting stents have been shown to markedly improve angiographic and clinical outcomes after percutaneous coronary revascularization, but their performance in the treatment of de novo coronary lesions has not been compared in a prospective multicenter study.

Objective To compare the safety and efficacy of sirolimus-eluting vs paclitaxel-eluting coronary stents.

Design Prospective, randomized comparative trial (the REALITY trial) conducted between August 2003 and February 2004, with angiographic follow-up at 8 months and clinical follow-up at 12 months.

Setting Ninety hospitals in Europe, Latin America, and Asia.

Patients A total of 1386 patients (mean age, 62.6 years; 73.1% men; 28.0% with diabetes) with angina pectoris and 1 or 2 de novo lesions (2.25-3.00 mm in diameter) in native coronary arteries.

Intervention Patients were randomly assigned in a 1:1 ratio to receive a sirolimus-eluting stent (n = 701) or a paclitaxel-eluting stent (n = 685).

Main Outcome Measures The primary end point was in-lesion binary restenosis (presence of a more than 50% luminal-diameter stenosis) at 8 months. Secondary end points included 1-year rates of target lesion and vessel revascularization and a composite end point of cardiac death, Q-wave or non–Q-wave myocardial infarction, coronary artery bypass graft surgery, or repeat target lesion revascularization.

Results In-lesion binary restenosis at 8 months occurred in 86 patients (9.6%) with a sirolimus-eluting stent vs 95 (11.1%) with a paclitaxel-eluting stent (relative risk [RR], 0.84; 95% confidence interval [CI], 0.61-1.17; P = .31). For sirolimus- vs paclitaxel-eluting stents, respectively, the mean (SD) in-stent late loss was 0.09 (0.43) mm vs 0.31 (0.44) mm (difference, −0.22 mm; 95% CI, −0.26 to −0.18 mm; P<.001), mean (SD) in-stent diameter stenosis was 23.1% (16.6%) vs 26.7% (15.8%) (difference, −3.60%; 95% CI, −5.12% to −2.08%; P<.001), and the number of major adverse cardiac events at 1 year was 73 (10.7%) vs 76 (11.4%) (RR, 0.94; 95% CI, 0.69-1.27; P = .73).

Conclusion In this trial comparing sirolimus- and paclitaxel-eluting coronary stents, there were no differences in the rates of binary restenosis or major adverse cardiac events.

Clinical Trial Registration ClinicalTrials.gov Identifier: NCT00235092

Figures in this Article

A report in June 2002 of a profoundly reduced 6-month restenosis rate among recipients of sirolimus-eluting stents for the treatment of de novo coronary lesions1 was the first of a rapidly growing number of studies showing the safety and efficacy of this stent in the management of coronary artery disease.25 Favorable clinical results were also reported with a paclitaxel-eluting stent, which uses a different drug, polymer, and drug-release kinetics.69 The results of randomized trials comparing each of these drug-eluting stents with bare metal stents of identical or similar design indicate that the rate of late luminal loss is lower after implantation of sirolimus-eluting stent than paclitaxel-eluting stent.10 However, differences in the populations studied, in implantation and angiographic techniques applied, and in methods of data analysis may not allow for legitimate comparisons of these different studies. Thus, a prospective, randomized comparison of sirolimus-eluting stent vs paclitaxel-eluting stent was warranted.

The objective of this study was to compare the safety and efficacy of the CYPHER sirolimus-eluting stent (Cordis Corp, Warren, NJ) and the TAXUS paclitaxel-eluting stent (Boston Scientific Corp, Natick, Mass) systems in a multicenter, randomized clinical trial in patients with de novo coronary artery lesions.

Patient Enrollment

This prospective trial randomized 1386 patients between August 2003 and February 2004 at 90 centers in Europe, Latin America, and Asia. The protocol was reviewed and approved by the ethics committee of each participating medical institution. Prior to any test or procedure related to the trial, the benefits and risks of the study were explained and written informed consent was obtained from each participating patient. The patients enrolled in this trial were aged 18 years or older, presented with 1 or 2 de novo lesions 2.25 mm to 3.00 mm in diameter by visual estimate in 1 or 2 native coronary arteries, and stable or unstable angina pectoris, using the Canadian Cardiology Society (I-IV) and Braunwald (B and C, I-III) classifications,11,12 or documented silent ischemia. One target lesion 15 mm or more and a second lesion 10 mm or more in length could be treated (without upper limit in lesion length), with visually estimated stenosis(es) between 51% and 99% and a Thrombolysis in Myocardial Infarction (TIMI) grade 1 or greater coronary flow designation for both lesions.

Patients were excluded from the trial if they (1) had a Q-wave or non–Q-wave myocardial infarction (MI) within 72 hours, with an initial creatine kinase level of more than twice the upper limit of normal and creatine kinase and creatine kinase–MB fraction persistently abnormal at the time of the procedure; (2) presented with Braunwald A I-II-III unstable angina; (3) had a more than 50% unprotected left main coronary stenosis or more than 50% stenoses of additional lesions proximal or distal to the target lesion; (4) had any target lesion containing a thrombus or calcifications that precluded successful predilatation or was totally occluded; (5) had a left ventricular ejection fraction of less than 25%; (6) had a serum creatinine level of more than 2.9 mg/dL (260 μmol/L) at the time of the procedure; or (7) had allergy to aspirin, clopidogrel, ticlopidine, heparin, stainless steel, contrast material, sirolimus, or paclitaxel. Contraindications to undergo coronary artery bypass graft surgery, pretreatment with methods other than balloon angioplasty, lesion tortuosity precluding proper stent delivery or deployment, prior stent implantation within 10 mm of the target lesion(s), previous brachytherapy, cardiac allograft, and life expectancy of less than 12 months were other possible reasons for exclusion from the trial.

Randomization Procedure

Following identification of a target lesion that met all eligibility criteria, patients received a unique study identification code and were randomly assigned on a 1:1 basis to receive 1 of the 2 study stents. Randomization was stratified according to participating site and number of lesions and concealed using a central telephone allocation service. All patients underwent protocol-mandated follow-up angiography at 8 months and were followed up clinically at 30 days and 8, 12, 18, and 24 months after the index procedure. This article reports the 8-month angiographic and 12-month clinical results (24-month follow-up is not yet available).

Procedural Techniques

Percutaneous vascular access was obtained according to each institution's standard procedures. After administration of nitrates, balloon predilatation of the target lesion was performed before delivery of 1 or more stents of sufficient length to completely cover the target lesions. The sirolimus-eluting stents were delivered on a Raptor Rapid Exchange (Cordis Corp) balloon catheter. The paclitaxel-eluting stents were delivered on a Maverick (Boston Scientific Corp) balloon catheter. The size of the sirolimus-eluting stent ranged between 8 mm and 33 mm in length and between 2.25 mm and 3.00 mm in diameter, while the size of the paclitaxel-eluting stent ranged between 8 mm and 32 mm in length and between 2.25 mm and 3.00 mm in diameter. Use of different drug-eluting stents in the same patient was not allowed.

Quantitative Coronary Angiography

An independent angiographic core laboratory (Cardialysis, Rotterdam, the Netherlands) analyzed all preprocedural, periprocedural, and postprocedural angiographic images using edge-detection techniques.13 The core laboratory was blinded to the treatment assignment. (The 2 types of stent have a similar angiographic appearance.) Coronary luminal diameter and degree of stenosis (as a percentage of the diameter) were measured before dilatation, at the end of the procedure, and at the 8-month angiographic follow-up. Binary restenosis was defined as the presence of a more than 50% luminal-diameter stenosis. Late loss was calculated as the difference between minimum luminal diameter (MLD) immediately after the procedure and MLD measured at 8 months. The target lesion was defined as the stent segment and 5 mm proximal and distal to the edge of the stent.

Periprocedural and Long-term Antithrombotic Regimen

The following guidelines were specified by the protocol regarding the administration of antithrombotic medications:

Preprocedure. Treatment with aspirin began 12 hours or more before the procedure in a dose of at least 100 mg. Clopidogrel was administered before or immediately after the procedure in a loading dose of 300 mg followed by 75 mg once daily or in a maintenance dose of 75 mg for 3 or more days before the procedure. Alternatively, 2 doses of ticlopidine, 250 mg, were administered within 24 hours before the revascularization procedure.

Intraprocedure. Heparin was administered in boluses to reach and maintain an activated clotting time of more than 250 seconds. The use of glycoprotein IIb/IIIa inhibitors was left to investigators' discretion.

Postprocedure. Clopidogrel, 75 mg, was administered once daily, or ticlopidine, 250 mg, was administered twice daily.

Long-term. Aspirin, 100 mg/d, was administered indefinitely to all patients. Clopidogrel, 75 mg once daily, was administered for 6 months or more in the paclitaxel-eluting stent group and for 2 months or more in the sirolimus-eluting stent group. Alternatively, ticlopidine, 250 mg twice daily, was administered for 6 months or more in the paclitaxel-eluting stent group and for 2 months or more in the sirolimus-eluting stent group.

Patient Follow-up

All surviving patients were to have repeat angiography at 8 months ± 30 days of follow-up. Clinical follow-up visits were scheduled at 30 days, 8 months, and 12 months and included a physical examination at 8 months and, at all other time points, monitoring of cardioactive and antithrombotic drug use, interim hospitalizations, invasive or noninvasive diagnostic tests, and occurrence of major adverse cardiac events as well as stable or unstable angina according to the Canadian Cardiology Society and the Braunwald classifications.11,12

Prespecified Study End Points

The primary end point of the trial was the rate of binary in-lesion restenosis by quantitative coronary angiography (QCA) at 8 months after the index procedure.

The secondary end points of the trial included rates of target lesion revascularization (TLR); target vessel revascularization; target vessel failure, defined as cardiac death, MI, or target vessel revascularization; composite major adverse cardiac events, including cardiac death, Q-wave or non–Q-wave MI, emergent coronary artery bypass graft surgery, or repeat TLR; in-stent binary restenosis by QCA; and in-stent and in-lesion late loss by QCA, up to 8 months of follow-up. Additional secondary angiographic and procedural end points included in-stent and in-lesion MLD and percentage diameter stenosis by QCA immediately after the index procedure and at 8 months of follow-up; device success, defined as attainment of a final residual diameter stenosis of less than 30% by QCA, using the assigned device only; lesion success, defined as the attainment of less than 50% residual stenosis by QCA, using any percutaneous revascularization method; and procedure success, defined as attainment of less than 50% final diameter stenosis by QCA, using any percutaneous revascularization method, without death, MI, or repeat TLR during the index hospitalization. There were no prespecified subgroup analyses. Target lesion revascularization was considered clinically driven if prompted by symptoms consistent with myocardial ischemia, preceded by an abnormal stress test result consistent with myocardial ischemia, if there were other electrocardiographic changes consistent with myocardial ischemia, or if the lesion diameter stenosis was more than 70% at follow-up.

Definition of Stent Thrombosis

Stent thrombosis was defined as a composite, 30-day end point including death, Q-wave MI, or abrupt vessel closure requiring revascularization. Any death not attributed to a noncardiac cause in the first 30 days or any Q-wave MI in the territory of the stented vessel in the first 30 days was adjudicated as stent thrombosis. It was classified as acute if it occurred within the first 24 hours, subacute up to 30 days, and late after 30 days. Late thrombosis was defined as MI attributable to the target vessel with angiographic documentation (site-reported or by QCA) of thrombus or total occlusion at the target site more than 30 days after the index procedure in absence of an interim target vessel revascularization.

Trial Monitoring

A data and safety monitoring board was responsible for the review of data and identification of potential safety issues. The members of this Board were not affiliated with the study sponsor and did not participate in the trial. Meetings of the board were planned regularly to allow the review of all reported major adverse cardiac events throughout the duration of the study. The data and safety monitoring board was unaware of the treatment assignment of individual patients.

A clinical events committee was composed of interventional and noninterventional cardiologists not associated with the sponsor. This committee developed specific criteria for the classification of major clinical events. The members of the committee were not provided with the treatment assignment of individual patients or the primary results of the trial. The committee met regularly to review and adjudicate all major clinical events, including stent thrombosis. All electrocardiograms related to clinical events and other pertinent data were adjudicated by the clinical events committee, who classified the MIs based on available data.

Statistical Analysis

The trial was designed with a 2-sided, P<.05 level of significance and 95% power to reject the null hypothesis of no difference between the 2 treatment groups. The assumptions used for the power calculations were a 14.0%, 8-month in-lesion restenosis rate with paclitaxel-eluting stent vs 8.0% with the sirolimus-eluting stent; ie, a 43% reduction in restenosis at 8 months with the sirolimus-eluting stent. The assumptions for the sirolimus-eluting stent were based on the results of the RAVEL,1 SIRIUS,2 and E-SIRIUS3 studies, and for the paclitaxel-eluting stent on the results of the TAXUS-II trial7 (the TAXUS-IV8 results were not yet available at the time of our study design). Applying these assumptions, a sample size of 1476 lesions was estimated for the trial under the further assumption that complete 8-month follow-up information would be available for each patient in the study. The total sample size was increased to 2000 lesions (1334 patients), to account for 74% compliance with the 8-month angiographic follow-up, and an average of 1.5 lesions per patient. From the 1386 patients randomly assigned initially, 1353 (n = 684 with a sirolimus-eluting stent and n = 669 with a paclitaxel-eluting stent) were included in the analysis (Figure 1).

Figure 1. Flow of Patients Through the Trial
Graphic Jump Location

The protocol prespecified a modified intention-to-treat principle whereby only randomized patients who underwent an attempt to implant the assigned study stent were included in the analysis; this led to the exclusion of 17 patients allocated to sirolimus-eluting stent and 16 patients allocated to paclitaxel-eluting stent (Figure 1). In a sensitivity analysis of clinical outcomes specified post hoc, we included 684 patients allocated to sirolimus-eluting stent and 669 patients allocated to paclitaxel-eluting stent. Three patients had to be excluded from this post hoc analysis; 1 in each group had erroneously undergone randomization without giving written informed consent, and another patient allocated to sirolimus-eluting stent withdrew consent immediately after randomization. Ascertainment of angiographic outcomes was impossible in patients not attending follow-up angiography. This resulted in the additional exclusion of 48 patients allocated to sirolimus-eluting stent and 61 patients allocated to paclitaxel-eluting stent in the analysis of angiographic outcomes (Figure 1). (For the clinical analysis, patients followed up included those who did not undergo 8-month angiographic follow-up.)

The correlation of lesion characteristics within patients with multiple lesions had negligible effects on standard errors: the design factor for differences in in-lesion binary restenosis, defined as the robust standard error adjusted for the correlation of multiple lesions within a patient divided by the conventional standard error assuming no correlation, was 1.05.

Clinical events including death, MI, and revascularization are reported on a per-patient basis. For patients with multiple lesions, a failure of any lesion was counted toward the composite event rate. Differences between the treatment groups were examined by analysis of variance for continuous variables and by the Fisher exact test for categorical variables. The incidence of major adverse cardiac events during the follow-up period was analyzed using actuarial life-table methods. All statistical analyses were performed using SAS statistical software, version 8 (SAS Institute Inc, Cary, NC).

There were 684 patients with 970 lesions assigned to treatment with sirolimus-eluting stent and 669 patients with 941 lesions assigned to paclitaxel-eluting stent. Baseline characteristics are presented in Table 1. The mean age of the overall population was 62.6 years, 989 (73.1%) were men, and diabetes mellitus was present in 379 (28.0%). There were 769 patients (50%) who presented with stable angina, 406 (30%) had unstable angina, and the remainder had silent ischemia. There were no significant differences in these baseline characteristics between the 2 study groups.

Table Graphic Jump LocationTable 1. Baseline Demographic and Clinical Characteristics*
Baseline Lesion Characteristics

Table 2 shows the lesion characteristics of the overall population and of each study group. There were no significant differences between the groups. Nearly 50% of the target lesions were located in the left anterior descending coronary artery, while nearly all other lesions were evenly distributed between the left circumflex and the right coronary arteries. The lesion length was more than 10 mm in three fourths and more than 20 mm in one fourth of lesions, and one third were moderately or heavily calcified.

Table Graphic Jump LocationTable 2. Baseline Lesion Characteristics
Procedural Characteristics and Outcomes

A single stent was implanted in 546 patients (40%), 2 stents in 487 (36%), 3 stents in 206 (15%), and 4 or more stents in 111 (8%) in both study groups (mean, 1.9 stents per patient and 1.4 stents per lesion). The mean stent diameter was 2.8 mm in both groups, and the mean length was 22.8 mm in the sirolimus-eluting stent group and 23.5 mm in the paclitaxel-eluting stent group. The maximum dilatation pressure during stent implantation was significantly lower in the paclitaxel-eluting stent group than the sirolimus-eluting stent group. Other procedural characteristics, including prestent angioplasty, total stent length, postprocedure TIMI grade, rates and types of postprocedure intimal dissection, and use of glycoprotein IIb/IIIa inhibitors during the procedure, were similar in both study groups (Table 3).

Procedural Success and Quantitative Coronary Angiography

There were no differences between the 2 study groups in mean lesion length, preprocedure reference vessel diameter, percentage diameter stenosis, or minimum luminal diameter. There was no significant difference in the delivery success of the 2 devices. More than 94% of lesions were successfully treated with the assigned stent in both study groups, and ultimately, less than 50% residual stenosis using any percutaneous revascularization method was achieved in nearly 100% of the stented lesions (Table 4).

Table Graphic Jump LocationTable 4. Results of Quantitative Coronary Analysis at Baseline and at 8-Month Follow-up*

The mean in-lesion binary restenosis rate, the primary study end point, was 9.6% (n = 86)in the sirolimus-eluting stent group vs 11.1% (n = 95) in the paclitaxel-eluting stent group, a difference that did not reach statistical significance (relative risk, 0.84; 95% confidence interval, 0.61-1.17; P = .31). However, significant differences were observed between the 2 groups in some immediate outcomes ascertained by QCA. In particular, the postprocedure mean in-stent minimum luminal diameter was significantly smaller, and in-stent percentage diameter stenosis was significantly greater in the sirolimus-eluting stent group than in the paclitaxel-eluting stent group. Despite this, at 8 months, QCA measurements, available for 1244 (91.9%) of 1353 patients and for 1754 (91.7%) of 1911 lesions, showed significant differences in MLD, absolute gain, late loss, and late loss index, all favoring the sirolimus-eluting stent (Table 4).

A sensitivity analysis of clinical outcomes based on 699 patients allocated to sirolimus-eluting stent and 684 patients allocated to paclitaxel-eluting stent yielded nearly identical results.

Short- and Long-term Adverse Clinical Events

Death, Q-wave or non–Q-wave MI, and surgical or percutaneous TLR occurred during hospitalization in 4 (0.6%), 29 (4.2%), and 3 (0.4%) patients, respectively, in the group assigned to sirolimus-eluting stent vs 2 (0.3%), 31 (4.6%), and 1 (0.1%) patients, respectively, in the group assigned to paclitaxel-eluting stent. The overall rate of in-hospital major adverse cardiac events was 4.5% in the sirolimus-eluting stent group vs 4.9% in the paclitaxel-eluting stent group. These differences were not statistically significant. Likewise, there were no significant differences between the 2 study groups in rates of other in-hospital adverse clinical events, including target vessel revascularization, stent thrombosis, cerebrovascular events, and hemorrhagic complications.

The overall, 12-month cumulative rate of major adverse cardiac events was 10.7% (n = 73) in the sirolimus-eluting stent group vs 11.4% (n = 76) in the paclitaxel-eluting stent group (Table 5 and Figure 2). The overall incidence of TLR was 6.0% (n = 41) in the sirolimus-eluting stent group vs 6.1% (n = 41) in the paclitaxel-eluting stent group. In the sirolimus-eluting stent group, 50.0% of all TLRs were clinically driven vs 54.3% of TLRs in the paclitaxel-eluting stent group. These differences were not statistically significant. Among other adverse clinical events, there were no differences between the 2 groups in rates of target vessel revascularization, target vessel failure, subacute occlusion, cerebral vascular accidents, or hemorrhagic complications. There were 5 stent thromboses (0.7%) in the sirolimus-eluting stent group vs 13 stent thromboses (1.9%) in the paclitaxel-eluting stent group, a difference that nearly reached statistical significance (P = .06 by Fisher exact test).

Table Graphic Jump LocationTable 5. Major Adverse Clinical Events During 12 Months of Follow-up*
Figure 2. Kaplan-Meier 12-Month Actuarial Incidence of Major Adverse Cardiac Events
Graphic Jump Location
Compliance With Antiplatelet Drug Regimen

The overall mean (SD) duration of antiplatelet therapy was 175 (74) days in the sirolimus-eluting stent group vs 204 (47) days in the paclitaxel-eluting stent group (P<.001). At the time of discharge from the hospital, 96.8% and 97.0% of patients in the sirolimus-eluting stent and paclitaxel-eluting stent groups, respectively, were treated with doses of clopidogrel or ticlopidine according to the protocol-mandated guidelines. Corresponding values were 95.9% and 97.6% of patients at 1 month and 48.8% and 52.5% of patients at 8 months of follow-up. These differences in treatment adherence rates between the 2 groups were not statistically significant.

The REALITY trial is the first large, randomized, multicenter trial to directly compare the clinical and angiographic results after percutaneous myocardial revascularization with a sirolimus-eluting stent vs a paclitaxel-eluting stent in a population of patients with de novo coronary artery stenoses. The baseline clinical and angiographic characteristics of the study population were consistent with those found in other interventional studies of angioplasty with or without stent implantation for single- or multivessel coronary artery disease outside of the acute phase of MI, including nearly 30% with diabetes. In addition, the mean reference vessel diameter in this study (2.40 mm) was smaller than in the SIRIUS2 (2.80 mm) or the TAXUS-IV8 (2.75 mm) trials.

Early Results

From a procedural standpoint, this study demonstrated a similarly high deliverability of both stents and similar immediate procedural outcomes in both study groups. However, the angiographic measurements made immediately after stent implantation suggested a slightly larger lumen of the paclitaxel-eluting stent than that of the sirolimus-eluting stent, despite the use of a higher maximum dilatation pressure with the latter, perhaps because of a greater compliance of the balloon used to deploy the TAXUS stent. As observed in earlier trials comparing each stent type with their bare metal counterparts, the rates of procedural and in-hospital complications were low and similar in both study groups.

Long-term Angiographic Results

At the 8-month angiographic follow-up, significant differences were found in in-stent MLD, percentage diameter stenosis, in-stent late loss, and in-stent late loss index, all favoring the sirolimus-eluting stent. These observations indicate a significantly greater degree of suppression of neointimal hyperplasia achieved by the sirolimus-eluting stent, particularly considering the smaller mean vessel diameter measured immediately after the implantation of sirolimus-eluting stent. However, the significant differences in several continuous angiographic variables, most importantly in-stent late loss, did not translate into significant differences in in-lesion binary restenosis or in TLR.

The discordance between late loss and binary restenosis at 8 months is probably multifactorial. First, immediately after stent deployment, a smaller MLD was achieved with the sirolimus-eluting stent than with the paclitaxel-eluting stent. Although the mean difference of 0.08 mm might seem trivial, it is of the same magnitude as the mean 0.09-mm total in-stent late loss observed with the sirolimus-eluting stent and might therefore have represented a notable handicap for this stent. Second, the relationship between late loss and binary restenosis is nonlinear and has been characterized by a curvilinear function.14 The observed late loss values for both sirolimus-eluting stent and paclitaxel-eluting stent are located on the left of the curve, near an x-axis value of 0, where its slope is shallow. Furthermore, binary restenosis is by definition a dichotomous variable, which is only declared when a threshold of 50% diameter stenosis is reached, below which there is no further discrimination. Third, the large experience gained with bare metal stenting has confirmed that differences in stent performance are more pronounced when the lesions are more complex and when the patients are at higher baseline risk of developing restenosis.15,16 Since patients with complex lesions and at higher periprocedural risk are generally less likely to be enrolled into randomized trials, those enrolled in the REALITY trial had only moderately complex lesions. Nearly 50% of the patients had single-vessel disease, the majority of lesions ranged between 10 mm and 20 mm in length, and the predominant lesion type was B2. These characteristics probably reduced the likelihood of identifying a difference in performance between the 2 stents.

Long-term Clinical Results

The 12-month rates of major adverse cardiac events were similarly low in both study groups, confirming the excellent performance of both stents. In particular, there was no statistically significant difference in any of the major clinical events, including death, MI, and TLR. A longer follow-up may be required for the difference in late loss to translate into a significant difference in event rates. It is conceivable that a 12-month follow-up was not sufficient to unmask the importance of suppression or delay of neointimal hyperplasia.

In contrast with the results of other studies,17,18 the rates of stent thrombosis in this study were similar in both groups. Since late stent thrombosis is a rare event, larger trials or meta-analyses will be needed to determine whether the trends observed in single studies or registries are the expression of a worrisome adverse clinical event.

Comparison With Similar Studies

The TAXi study randomly assigned 202 consecutive patients to the implantation of sirolimus-eluting stent vs paclitaxel-eluting stent.19 At 6 months, there was no difference in major adverse cardiac events, defined as death, MI, and TLR, though the study was prematurely terminated and limited by a small patient population. More recently, the results of ISAR-DESIRE,20 ISAR-DIABETES,21 and SIRTAX,22 which directly compared sirolimus-eluting stent with paclitaxel-eluting stent, were presented. Unlike the REALITY trial, these studies showed a clear and statistically significant superiority of sirolimus-eluting stent in the primary end points; namely, in-segment restenosis, in-segment late loss, and the composite of death, MI, and TLR, respectively. Importantly, these trials differ from REALITY by having enrolled patients with higher lesion complexity (exclusively in-stent restenosis in ISAR-DESIRE) or patients with specific and prorestenotic comorbidity (exclusively patients with diabetes in ISAR-DIABETES) or by an enrollment unlimited by exclusion criteria (SIRTAX). Each study was conducted in no more than 5 medical centers, in contrast with the 90 worldwide centers that participated in REALITY. The lower risk profile and lower risk of restenosis among the patients enrolled in the REALITY trial, together with an inevitable variability in interventional techniques, may have contributed to the smaller-than-expected difference in the primary end point.

A recently published meta-analysis of the results of 6 completed clinical trials, in which a total of 3669 patients were randomly assigned to sirolimus-eluting stent vs paclitaxel-eluting stent, found significant differences in rates of restenosis (9.3% vs 13.1%; P<.001) and TLR (5.1% vs 7.8%; P<.001) favoring the sirolimus-eluting stent.23 No significant differences were found in mortality, MI, or stent thrombosis between the study groups.

Limitations

As is the case with nearly all randomized trials, several enrollment exclusion criteria created a selected patient population. Consequently, the observations made in the REALITY trial apply mostly to patients treated for 1 or 2 de novo lesions in small to moderately sized coronary arteries. Other ongoing trials that enroll mostly patients with diabetes, patients with in-stent restenosis or bifurcated or very long lesions, or patients with acute MI will help define the role of sirolimus-eluting stent vs paclitaxel-eluting stent in the overall percutaneous management of coronary artery disease. The respective contributions of these studies and of the REALITY trial should be viewed as complementary and not as competitive or necessarily discordant.

In this multicenter randomized comparison of sirolimus-eluting stent vs paclitaxel-eluting stent in patients with 1 or 2 de novo lesions in relatively small coronary arteries, no statistically significant difference was observed in the rate of binary in-lesion restenosis at 8 months, although QCA measurements indicated a more profound inhibition of neointimal hyperplasia by sirolimus-eluting stent. A longer follow-up may be required for the different degrees of neointimal proliferation suppression to translate into significantly different rates of binary restenosis or adverse clinical events.

Corresponding Author: Marie-Claude Morice, MD, Institut Cardiovasculaire Paris Sud, Institut Hospitalier Jacques Cartier, Avenue du Noyer Lambert, F-91300 Massy, France (mc.morice@icps.com.fr).

Author Contributions: Dr Morice had full access to all of the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.

Study concept and design: Morice, Colombo.

Acquisition of data: Morice, Colombo, Tamburino, Guagliumi, Sousa.

Analysis and interpretation of data: Morice, Colombo, Meier, Serruys, Stoll.

Drafting of the manuscript: Morice, Colombo, Stoll.

Critical revision of the manuscript for important intellectual content: Morice, Colombo, Meier, Serruys, Tamburino, Guagliumi, Sousa, Stoll.

Study supervision: Morice, Stoll.

Financial Disclosures: Dr Meier has received research support from both Cordis and Boston Scientific.

Funding/Support: This study was sponsored by Cordis Corporation (a Johnson & Johnson Company) Clinical Research Europe, Waterloo, Belgium.

Role of the Sponsor: The sponsor designed and monitored the study and assisted in the preparation of the manuscript. Approval of the sponsor was not required prior to manuscript submission. Data management was performed independently from the sponsor by Cardialysis, an external clinical research organization.
Gerrit-Anne Van Es, PhD, director of Research and Development for Cardialysis, Rotterdam, the Netherlands, an independent clinical research and data management organization, and Eric van Remortel, an employee of Cardialysis, maintained the entire trial database, completed an independent statistical review of every data point, and provided reports to the sponsor. Cardialysis received compensation for data management and statistical analysis, but Dr Van Es did not receive personal compensation for his role in this trial.

Independent Statistical Review: Christophe E. Minder, PhD, professor of Medical Statistics at the University of Bern, Switzerland, and Peter Jüni, MD, senior lecturer in Clinical Epidemiology at the University of Bern, received a complete copy of the raw data from Cordis Corporation and performed an independent statistical analysis. They received no compensation for this work and had no conflicts of interest, not receiving any type of payment, equity, or reimbursement from either of the companies manufacturing the stents compared in the trial. They confirmed that they were able to replicate the analyses of the primary angiographic and secondary clinical outcomes reported in the manuscript and that they consider the analyses to be appropriate.

The REALITY Trial Investigators: steering committee: Antonio Colombo, MD, Bernhard Meier, MD, Patrick Serruys, MD, PhD; clinical events committee: Aad van den Bos, MD, Jeroen Vos, MD, Claude Hanet, MD, Eelko Ronner, MD. Peter de Jaegere, MD; data and safety monitoring board chairman: Jan G. P. Tijssen, PhD, Freek W. A. Verheugt, MD, PhD, William Wijns, MD, PhD; participating investigators and institutions: principal investigator: Marie-Claude Morice, MD, Institut Cardiovasculaire Paris Sud, Massy, France; coinvestigators: Belgium: Luc Janssens, MD, Imelda Hospital, Bonheiden; Frank Van Den Branden, MD, University Hospital Middelheim, Antwerp; Victor Legrand, MD, CHU Sart Tilman, Liège; Walter Desmet, MD, K. U. Leuven, Leuven; the Netherlands: Patrick W. Serruys, MD, Erasmus Medical Center Rotterdam, Rotterdam; Hans Bonnier, MD, Catharina Hospital, Eindhoven; Maarten Jan Suttorp, MD, Sint Antonius Hospital, Nieuwegein; Felix Zijlstra, MD, University Hospital Groningen, Groningen; France: Jean Fajadet, MD, Clinique Pasteur, Toulouse; Khalifé Khalife, MD, CHR Bon Secours, Metz; Emmanuel Teiger, MD, Hôpital Henri Mondor, Creteil; Henri Escojido, MD, Centre Hospitalier Prive Clairval, Marseille; Hélène Eltchaninoff, MD, and Alain Cribier, MD, CHRU Rouen Hôpital Charles-Nicolle, Rouen; Gabriel Steg, MD, CHU Bichat-Claude Bernard, Paris; Philippe Guyon, MD, Centre Cardiologique Du Nord, Saint Denis; Didier Carrié, MD, Hopital Rangueil, Toulouse; Paul Barragan, MD, Centre Hospitalier Privé Beauregard, Marseille; Jean-Pierre Bassand, MD, Centre Hospitalier Universitaire Jean Minjoz, Besançon; Philippe Commeau, MD, Centre Hospitalier Privé Beauregard, Marseille; Christian Spaulding, MD, Hôpital Cochin, Paris; Philippe Garot, MD, Centre Hospitalier Privé Claude Galien, Quincy Sous Senart; Olivier Varenne, MD, Hôpital Cochin, Paris; Germany: Wolfgang Rutsch, MD, Univ. Klin. Charité Der Humboldt University, Berlin; Joachim Schofer, MD, Kardiologische Gemeinschaftspraxis, Hamburg; Christoph Nienaber, MD, Universitatsklinikum Rostock, Rostock; Christof Bode, MD, Albert-Ludwigs Universität Freiburg, Freiburg; Hans Störger, MD, Rotes Kreuz Krankenhaus; Frankfurt; Gerhard Schuler, MD, Herzzentrum Leipzig, Leipzig; Eberhard Grube, MD, Krankenhaus Siegburg Gmbh, Siegburg; Johannes Brachmann, MD, Klinikum Coburg Gmbh, Coburg; Dieter Horstkotte, MD, Herz-und Diabeteszentrum Nordrhein-Westfalen, Bad Oeynhausen; Karl-Heinz Kuck, MD, Allgemeines Krankenhaus St. Georg II, Hamburg; Harald Mudra, MD, Städtisches Krankenhaus, Neuperlach, Munich; Andreas Mügge, MD, Ruhr Universität Bochum St. Josef Krankenhaus, Bochum; Andreas M. Zeiher, MD, Klin. Der J. W. Goethe Universität, Frankfurt; Italy: Antonio Colombo, MD, Fondazione S. Raffaele/Centro Cuore, Milan; Bernhard Reimers, MD, Ospedale Di Mirano, Mirano; Ezio Bramucci, MD, Policlinico San Matteo, Pavia; Imad Sheiban, MD, Azienda Ospedaliera San Giovanni Battista Di Torino University, Turin; Antonio Bartorelli, MD, Centro Cardiologico “Ford. Monzino,” Milan; Giulio Guagliumi, MD, Ospedale Riuniti Di Bergamo, Bergamo; Corrado Tamburino, MD, Azienda Ospedaliera Vittorio Emanuele, Catania; United Kingdom: Keith D. Dawkins, MD, Southampton University Hospital Wessex, Southampton; Carlo Di Mario, MD, Royal Brompton Hospital, London; Keith G. Oldroyd, MD, Western Infirmary, Glasgow; Mark A. De Belder, MD, James Cook University Hospital, Middlesborough; David Thomas, MD, King's College Hospital, London; Rodney Stables, MD, Cardiothoracic Centre–Liverpool; Richard David Levy, MD, South Manchester University Hospital, Manchester; Peter Crean, MD, St James's Hospital, Dublin; David P. Foley, MD, Beaumont Hospital, Dublin; Declan Sugrue, MD, Mater Misericordiae Hospital Public, Dublin; Farzin-Fath Ordoubadi, MD, Manchester Heart Centre, Manchester; Stephen Holmberg, MD, Royal Sussex County Hospital, Brighton; Austria: Dietmar Glogar, MD, Allgemeines Krankenhaus Der Stadt Wien, Vienna; Sweden: Göran Olivecrona, MD, Lunds University Sjukhus, Lund; Denmark: Leif Thuesen, MD, Skejby Sygehus, Aarhus; Henning Kelbaek, MD, Heart Center, Rigshospitalet, Copenhagen; Jan Ravkilde, MD, Aalborg University Hospital, Aalborg; Per Thayssen, MD, Odense University Hospital, Odense; Ulrik Abildgaard, MD, Gentofte University Hospital, Hellerup; Norway: Jan Erik Nordrehaug, MD, Haukeland Sykehus, Bergen; Poland: Robert Gil, MD, Woloska MSWA Warszawa, Warsaw; Janusz Rzezniczak, MD, Instytut Kardiologii, Poznan; Dariusz Dudek, MD, Jagiellonian University, Krakow; Portugal: Ricardo Seabra-Gomes, MD, Hospital De La Santa Cruz, Linda-A-Velha; Jorge Quininha, MD, Hospital Santa Marta, Lisbon; José Baptista, MD, Unidade De Intervencao Cardiovascular (UIC), Portimao-Alvor; Serbia and Montenegro: Miodrag Ostojic, MD, Serbian Clinical Centre University, Belgrade; Switzerland: Stephan Windecker, MD, University Hospital, Bern; Spain: Carlos Macaya, MD, Hospital Clínico San Carlos, Madrid; Amadeo Betriu, MD, Hospital Clínico y Provincial, Barcelona; Thierry Colman, MD, Hospital Marqués De Valdecilla, Santander; Argentina: Jorge A. Belardi, MD, Instituto Cardiovascular De Buenos Aires, Buenos Aires; Liliana Grinfeld, MD, Hospital Italiano De Buenos Aires, Buenos Aires; Oscar Mendiz, MD, Institute of Cardiology & Cardiovascular Surgery, Buenos Aires; Czech Republic: Michail Zelizko, MD, Institute for Clinical & Experimental Medicine, Prague; Ladislav Groch, MD, Fakultni Nemocnice U Svate Anny, Brno; Greece: Vassilis Voudris, MD, Onassis Cardiac Surgery Center, Athens/Kalithea; Colombia: Juan Jose Arango, MD, Fundacion Valle De Lili, Cali; Brazil: Marco Perin, MD, Instituto Albert Einstein, Sao Paulo; J. Eduardo Sousa, MD, Instituto Dante Pazzanese, Sao Paulo; Mexico: Jorge Gaspar, MD, Instituto Nacional De Cardiologia “Ignacio Chavez,” Mexico City; Singapore: Philip Wong, MD, Singapore General Hospital; Tan Huay Cheem, MD, National University Hospital, Singapore; Hong Kong: Steven Li, MD, Queen Elizabeth Hospital, Hong Kong; Shu Kin Li, MD, Pamela Youde Nethersole Hospital, Hong Kong; India: Ashwinb Mehta, MD, Jaslok Hospital, Mumbai; K. Samuel Mathews, MD, Leelavati Hospital, Mumbai.

Acknowledgment: We thank Ingrid Pansar, MS, for her major contributions in managing this project, Dennis Donohoe, MD, for his critical review of the manuscript, and Gerrit-Anne Van Es, PhD, for providing an independent statistical analysis of the data. Ms Pansar and Dr Donohoe are employees of the sponsor and did not receive any compensation for their roles in this trial in particular.

Morice MC, Serruys PW, Sousa JE.  et al. RAVEL Study Group.  A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization.  N Engl J Med. 2002;346:1773-1780
PubMed   |  Link to Article
Moses JW, Leon MB, Popma JJ.  et al. SIRIUS Investigators.  Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery.  N Engl J Med. 2003;349:1315-1323
PubMed   |  Link to Article
Schofer J, Schluter M, Gershlick AH.  et al. E-SIRIUS Investigators.  Sirolimus-eluting stents for treatment of patients with long atherosclerotic lesions in small coronary arteries: double-blind, randomised controlled trial (E-SIRIUS).  Lancet. 2003;362:1093-1099
PubMed   |  Link to Article
Schampaert E, Cohen EA, Schluter M.  et al. C-SIRIUS Investigators.  The Canadian study of the sirolimus-eluting stent in the treatment of patients with long de novo lesions in small native coronary arteries (C-SIRIUS).  J Am Coll Cardiol. 2004;43:1110-1115
PubMed   |  Link to Article
Fajadet J, Morice MC, Bode C.  et al.  Maintenance of long-term clinical benefit with sirolimus-eluting coronary stents: three-year results of the RAVEL trial.  Circulation. 2005;111:1040-1044
PubMed   |  Link to Article
Grube E, Silber S, Hauptmann KE.  et al.  TAXUS I: six- and twelve-month results from a randomized, double-blind trial on a slow-release paclitaxel-eluting stent for de novo coronary lesions.  Circulation. 2003;107:38-42
PubMed   |  Link to Article
Colombo A, Drzewiecki J, Banning A.  et al.  Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions.  Circulation. 2003;108:788-794
PubMed   |  Link to Article
Stone GW, Ellis SG, Cox DA.  et al. TAXUS-IV Investigators.  One-year clinical results with the slow-release, polymer-based, paclitaxel-eluting TAXUS stent: the TAXUS-IV trial.  Circulation. 2004;109:1942-1947
PubMed   |  Link to Article
Stone GW, Ellis SG, Cox DA.  et al. TAXUS-IV Investigators.  A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease.  N Engl J Med. 2004;350:221-231
PubMed   |  Link to Article
Perin EC. Choosing a drug-eluting stent: a comparison between CYPHER and TAXUS.  Rev Cardiovasc Med. 2005;6:(suppl 1)  S13-S21
PubMed
Campeau L. Grading for angina pectoris [letter].  Circulation. 1976;54:522-523
PubMed
Braunwald E. Unstable angina: a classification.  Circulation. 1989;80:410-414
PubMed   |  Link to Article
Serruys PW, Foley DP, de Feyter PJQuantitative Coronary Angiography in Clinical Practice. Dordrecht, the Netherlands: Kluwer Academic; 1994
Mauri L, Orav EJ, O’Malley J.  et al.  Relationship of late loss in lumen diameter to coronary restenosis in sirolimus-eluting stents.  Circulation. 2005;111:321-327
PubMed   |  Link to Article
Edelman ER, Rogers C. Stent-versus-stent equivalency trials: are some stents more equal than others?  Circulation. 1999;100:896-898
PubMed   |  Link to Article
Hausleiter J, Kastrati A, Mehilli J.  et al.  Impact of lesion complexity on the capacity of a trial to detect differences in stent performance: results from the ISAR-STEREO trial.  Am Heart J. 2003;146:882-886
PubMed   |  Link to Article
Iakovou I, Schmidt T, Bonizzoni E.  et al.  Incidence, predictors, and outcome of thrombosis after successful implantation of drug-eluting stents.  JAMA. 2005;293:2126-2130
PubMed   |  Link to Article
Ong AT, Serruys PW, Aoki J.  et al.  The unrestricted use of paclitaxel- versus sirolimus-eluting stents for coronary artery disease in an unselected population: one-year results of the Taxus-Stent Evaluated at Rotterdam Cardiology Hospital (T-SEARCH) Registry.  J Am Coll Cardiol. 2005;45:1135-1141
PubMed   |  Link to Article
Goy JJ, Stauffer JC, Siegenthaler M, Benoit A, Seydoux C. A prospective randomized comparison between paclitaxel and sirolimus stents in the real world of interventional cardiology: the TAXi trial.  J Am Coll Cardiol. 2005;45:308-311
PubMed   |  Link to Article
Kastrati A, Mehilli J, von Beckerath N.  et al. ISAR-DESIRE Study Investigators.  Sirolimus-eluting stent or paclitaxel-eluting stent versus balloon angioplasty for prevention of recurrences in patients with coronary in-stent restenosis: a randomized controlled trial.  JAMA. 2005;293:165-171
PubMed   |  Link to Article
Kastrati A. Paclitaxel-eluting stent versus sirolimus-eluting stent for the prevention of restenosis in diabetic patients with coronary artery disease (ISAR-DIABETES). Paper presented at: 2005 Scientific Session of the American College of Cardiology; March 6, 2005; Orlando, Fla
Windecker S, Remondino A, Wenaweser P.  et al.  A randomized comparison of a sirolimus with a paclitaxel eluting stent for coronary revascularization: the SIRTAX trial [abstract].  J Am Coll Cardiol. 2005;45:(suppl A)  7A
Kastrati A, Dibra A, Eberle S.  et al.  Sirolimus-eluting stents vs paclitaxel-eluting stents in patients with coronary artery disease: meta-analysis of randomized trials.  JAMA. 2005;294:819-825
PubMed   |  Link to Article

Figures

Figure 1. Flow of Patients Through the Trial
Graphic Jump Location
Figure 2. Kaplan-Meier 12-Month Actuarial Incidence of Major Adverse Cardiac Events
Graphic Jump Location

Tables

Table Graphic Jump LocationTable 1. Baseline Demographic and Clinical Characteristics*
Table Graphic Jump LocationTable 2. Baseline Lesion Characteristics
Table Graphic Jump LocationTable 4. Results of Quantitative Coronary Analysis at Baseline and at 8-Month Follow-up*
Table Graphic Jump LocationTable 5. Major Adverse Clinical Events During 12 Months of Follow-up*

References

Morice MC, Serruys PW, Sousa JE.  et al. RAVEL Study Group.  A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization.  N Engl J Med. 2002;346:1773-1780
PubMed   |  Link to Article
Moses JW, Leon MB, Popma JJ.  et al. SIRIUS Investigators.  Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery.  N Engl J Med. 2003;349:1315-1323
PubMed   |  Link to Article
Schofer J, Schluter M, Gershlick AH.  et al. E-SIRIUS Investigators.  Sirolimus-eluting stents for treatment of patients with long atherosclerotic lesions in small coronary arteries: double-blind, randomised controlled trial (E-SIRIUS).  Lancet. 2003;362:1093-1099
PubMed   |  Link to Article
Schampaert E, Cohen EA, Schluter M.  et al. C-SIRIUS Investigators.  The Canadian study of the sirolimus-eluting stent in the treatment of patients with long de novo lesions in small native coronary arteries (C-SIRIUS).  J Am Coll Cardiol. 2004;43:1110-1115
PubMed   |  Link to Article
Fajadet J, Morice MC, Bode C.  et al.  Maintenance of long-term clinical benefit with sirolimus-eluting coronary stents: three-year results of the RAVEL trial.  Circulation. 2005;111:1040-1044
PubMed   |  Link to Article
Grube E, Silber S, Hauptmann KE.  et al.  TAXUS I: six- and twelve-month results from a randomized, double-blind trial on a slow-release paclitaxel-eluting stent for de novo coronary lesions.  Circulation. 2003;107:38-42
PubMed   |  Link to Article
Colombo A, Drzewiecki J, Banning A.  et al.  Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions.  Circulation. 2003;108:788-794
PubMed   |  Link to Article
Stone GW, Ellis SG, Cox DA.  et al. TAXUS-IV Investigators.  One-year clinical results with the slow-release, polymer-based, paclitaxel-eluting TAXUS stent: the TAXUS-IV trial.  Circulation. 2004;109:1942-1947
PubMed   |  Link to Article
Stone GW, Ellis SG, Cox DA.  et al. TAXUS-IV Investigators.  A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease.  N Engl J Med. 2004;350:221-231
PubMed   |  Link to Article
Perin EC. Choosing a drug-eluting stent: a comparison between CYPHER and TAXUS.  Rev Cardiovasc Med. 2005;6:(suppl 1)  S13-S21
PubMed
Campeau L. Grading for angina pectoris [letter].  Circulation. 1976;54:522-523
PubMed
Braunwald E. Unstable angina: a classification.  Circulation. 1989;80:410-414
PubMed   |  Link to Article
Serruys PW, Foley DP, de Feyter PJQuantitative Coronary Angiography in Clinical Practice. Dordrecht, the Netherlands: Kluwer Academic; 1994
Mauri L, Orav EJ, O’Malley J.  et al.  Relationship of late loss in lumen diameter to coronary restenosis in sirolimus-eluting stents.  Circulation. 2005;111:321-327
PubMed   |  Link to Article
Edelman ER, Rogers C. Stent-versus-stent equivalency trials: are some stents more equal than others?  Circulation. 1999;100:896-898
PubMed   |  Link to Article
Hausleiter J, Kastrati A, Mehilli J.  et al.  Impact of lesion complexity on the capacity of a trial to detect differences in stent performance: results from the ISAR-STEREO trial.  Am Heart J. 2003;146:882-886
PubMed   |  Link to Article
Iakovou I, Schmidt T, Bonizzoni E.  et al.  Incidence, predictors, and outcome of thrombosis after successful implantation of drug-eluting stents.  JAMA. 2005;293:2126-2130
PubMed   |  Link to Article
Ong AT, Serruys PW, Aoki J.  et al.  The unrestricted use of paclitaxel- versus sirolimus-eluting stents for coronary artery disease in an unselected population: one-year results of the Taxus-Stent Evaluated at Rotterdam Cardiology Hospital (T-SEARCH) Registry.  J Am Coll Cardiol. 2005;45:1135-1141
PubMed   |  Link to Article
Goy JJ, Stauffer JC, Siegenthaler M, Benoit A, Seydoux C. A prospective randomized comparison between paclitaxel and sirolimus stents in the real world of interventional cardiology: the TAXi trial.  J Am Coll Cardiol. 2005;45:308-311
PubMed   |  Link to Article
Kastrati A, Mehilli J, von Beckerath N.  et al. ISAR-DESIRE Study Investigators.  Sirolimus-eluting stent or paclitaxel-eluting stent versus balloon angioplasty for prevention of recurrences in patients with coronary in-stent restenosis: a randomized controlled trial.  JAMA. 2005;293:165-171
PubMed   |  Link to Article
Kastrati A. Paclitaxel-eluting stent versus sirolimus-eluting stent for the prevention of restenosis in diabetic patients with coronary artery disease (ISAR-DIABETES). Paper presented at: 2005 Scientific Session of the American College of Cardiology; March 6, 2005; Orlando, Fla
Windecker S, Remondino A, Wenaweser P.  et al.  A randomized comparison of a sirolimus with a paclitaxel eluting stent for coronary revascularization: the SIRTAX trial [abstract].  J Am Coll Cardiol. 2005;45:(suppl A)  7A
Kastrati A, Dibra A, Eberle S.  et al.  Sirolimus-eluting stents vs paclitaxel-eluting stents in patients with coronary artery disease: meta-analysis of randomized trials.  JAMA. 2005;294:819-825
PubMed   |  Link to Article

Letters

CME
Meets CME requirements for:
Browse CME for all U.S. States
Accreditation Information
The American Medical Association is accredited by the Accreditation Council for Continuing Medical Education to provide continuing medical education for physicians. The AMA designates this journal-based CME activity for a maximum of 1 AMA PRA Category 1 CreditTM per course. Physicians should claim only the credit commensurate with the extent of their participation in the activity. Physicians who complete the CME course and score at least 80% correct on the quiz are eligible for AMA PRA Category 1 CreditTM.
Note: You must get at least of the answers correct to pass this quiz.
You have not filled in all the answers to complete this quiz
The following questions were not answered:
Sorry, you have unsuccessfully completed this CME quiz with a score of
The following questions were not answered correctly:
Commitment to Change (optional):
Indicate what change(s) you will implement in your practice, if any, based on this CME course.
Your quiz results:
The filled radio buttons indicate your responses. The preferred responses are highlighted
For CME Course: A Proposed Model for Initial Assessment and Management of Acute Heart Failure Syndromes
Indicate what changes(s) you will implement in your practice, if any, based on this CME course.

Multimedia

Some tools below are only available to our subscribers or users with an online account.

Web of Science® Times Cited: 291

Related Content

Customize your page view by dragging & repositioning the boxes below.

See Also...
Articles Related By Topic
Related Collections
PubMed Articles